Translation of an existing implantable cardiac monitoring device for measurement of gastric electrical slow-wave activity.
Аннотация
BACKGROUND
Despite evidence that slow-wave dysrhythmia in the stomach is associated with clinical conditions such as gastroparesis and functional dyspepsia, there is still no widely available device for long-term monitoring of gastric electrical signals. Actionable biomarkers of gastrointestinal health are critically needed, and an implantable slow-wave monitoring device could aid in the establishment of causal relationships between symptoms and gastric electrophysiology. Recent developments in the area of wireless implantable gastric monitors demonstrate potential, but additional work and validation are required before this potential can be realized.
METHODS
We hypothesized that translating an existing implantable cardiac monitoring device, the Reveal LINQ™ (Medtronic), would present a more immediate solution. Following ethical approval and laparotomy in anesthetized pigs (n = 7), a Reveal LINQ was placed on the serosal surface of the stomach, immediately adjacent to a validated flexible-printed-circuit (FPC) electrical mapping array. Data were recorded for periods of 7.5 min, and the resultant signal characteristics from the FPC array and Reveal LINQ were compared.
KEY
RESULTS
The Reveal LINQ device recorded slow waves in 6/7 subjects with a comparable period (p = 0.69), signal-to-noise ratio (p = 0.58), and downstroke width (p = 0.98) to the FPC, but with reduced amplitude (p = 0.024). Qualitatively, the Reveal LINQ slow-wave signal lacked the prolonged repolarization phase present in the FPC signals.
CONCLUSIONS & INFERENCES: These findings suggest that existing cardiac monitors may offer an efficient solution for the long-term monitoring of slow waves. Translation toward implantation now awaits.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.