OBJECTIVE

This study developed a feedback-enabled magnetic device for the management of blepharoptosis and conducted a clinical trial to evaluate its efficacy and safety.

STUDY

DESIGN

Prospective, two-center, single-arm, open-label clinical trial.

SETTING

Tertiary referral centers.

METHODS

A device primarily comprising an optical sensor, a permanent electromagnetic actuator, and an iron sheet was developed. The palpebral fissure height (PFH) was measured, and the corrective gain of the paralytic upper eyelid was calculated. Efficacy was evaluated through symmetrical blinks, the corrective gain, and the electrical activity of the frontalis muscle. Safety was evaluated through subjective sensations and complications.

RESULTS

A total of 16 volunteers were recruited. After device-assisted correction, the patients reported subjective improvement of peripheral vision, and the mean PFH increased from 3.4 ± 1.6 to 7.2 ± 1.4 mm, with a mean gain of 3.7 ± 0.9 mm (range: 2.0-5.0 mm). The device could lift the paralytic eyelid without causing severe discomfort or complications. It also showed effects on the electrical activity of the frontalis muscle. Over the periorbital skin, posttest erythema was observed in one patient, mild hot sensation in two, and mild warm sensation in five. Subjective and objective evaluation of six patients who underwent two trials revealed no significant difference between the first and second trials (P > .05 for all seven items).

CONCLUSION

The concepts used in this device may provide the basis of a future device that may be of use in patients with blepharoptosis who are not suitable candidates for surgical correction. TRIAL REGISTERED AT CLINICALTRIALS.

GOV PRS: Registration number: NCT03812016.

Для цитирования: Lien Ching-Feng, Lin Bor-Shyh, Wu Tzong-Ming, Lin Yen-Chun, Chang Richard Chih-Hau, Cheng Cheng-Kuo, Yeh Jiann-Horng, Jeng Seng-Feng. Efficacy and Safety of a Magnetic Device for Temporary Management of Blepharoptosis: Single-Arm Open-Label Trial.. Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery. 2025. DOI: 10.1002/ohn.1350.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.