DOI 10.1089/end.2024.0315

Медицинская библиотека · Journal of endourology · 2024

Development and Initial Evaluation of a Cost-Effective Force Sensor for Ureteroscopic Application.

Gao Bruce M, Tsai Jacob C, Cumpanas Andrei D, Altamirano-Villarroel Jaime, Saadat Seyedamirvala, Pham Victor, Grohs Evan, Wu Yi Xi, Tano Zachary E, Ali Sohrab N, Jiang Pengbo, Patel Roshan M, Landman Jaime, Clayman Ralph V

СтатьяОценочное исследованиеПеревод пока не подготовлен

Аннотация

Purpose: Retrograde intrarenal surgery is the gold-standard treatment for most kidney stones. During ureteroscopy, ureteral access sheath insertion at forces greater than 8.0 Newtons (N) risks high-grade ureteral injury. To monitor force, our institution utilizes a unique, Bluetooth-equipped device (i.e., the University of California-Irvine Force Sensor). Given the unique nature of the force sensor, we sought to develop an inexpensive and accessible force sensor based on Boyle's law and the specific amount of force required to compress an occluded 1.0 mL syringe. Materials and

Methods

We evaluated three brands of 1.0 mL syringes. After setting the plunger at 1.0 mL, the syringe was occluded, and the syringe plunger was compressed. The syringe volume was recorded when the applied force on the plunger reached 4.0 N, 6.0 N, and 8.0 N. Multiple trials were performed to assess reliability and reproducibility. A method for applying this clinically was also developed.

Results

The precise force thresholds identified for a 1.0 mL Luer-Lok™ Syringe (Becton Dickinson, Franklin Lakes, NJ) were 0.30 mL for 4.00 N, 0.20 mL for 6.00 N, and 0.15 mL for 8.00 N. The 1.0 mL Tuberculin Syringe and 1.0 mL Luer Slip Syringe were less precise, but compression from 1.0 to 0.40 mL, 0.25 mL, and 0.20 mL corresponded to force sensor readings that did not exceed 4.00 N, 6.00 N, and 8.00 N, respectively.

Conclusions

Based on volume changes, 4.00 N, 6.00 N, and 8.00 N of force can be reliably and reproducibly achieved using an occluded 1.0 mL syringe.

Для цитирования: Gao Bruce M, Tsai Jacob C, Cumpanas Andrei D, Altamirano-Villarroel Jaime, Saadat Seyedamirvala, Pham Victor, Grohs Evan, Wu Yi Xi, Tano Zachary E, Ali Sohrab N, Jiang Pengbo, Patel Roshan M, Landman Jaime, Clayman Ralph V. Development and Initial Evaluation of a Cost-Effective Force Sensor for Ureteroscopic Application.. Journal of endourology. 2024. DOI: 10.1089/end.2024.0315.