DOI 10.1161/circep.122.010857

Медицинская библиотека · Circulation. Arrhythmia and electrophysiology · 2022

Forward-Solution Noninvasive Computational Arrhythmia Mapping: The VMAP Study.

Krummen David E, Villongco Christopher T, Ho Gordon, Schricker Amir A, Field Michael E, Sung Kevin, Kacena Katherine A, Martinson Melissa S, Hoffmayer Kurt S, Hsu Jonathan C, Raissi Farshad, Feld Gregory K, McCulloch Andrew D, Han Frederick T

СтатьяМногоцентровое исследованиеПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

The accuracy of noninvasive arrhythmia source localization using a forward-solution computational mapping system has not yet been evaluated in blinded, multicenter analysis. This study tested the hypothesis that a computational mapping system incorporating a comprehensive arrhythmia simulation library would provide accurate localization of the site-of-origin for atrial and ventricular arrhythmias and pacing using 12-lead ECG data when compared with the gold standard of invasive electrophysiology study and ablation.

METHODS

The VMAP study (Vectorcardiographic Mapping of Arrhythmogenic Probability) was a blinded, multicenter evaluation with final data analysis performed by an independent core laboratory. Eligible episodes included atrial and ventricular: tachycardia, fibrillation, pacing, premature atrial and ventricular complexes, and orthodromic atrioventricular reentrant tachycardia. Mapping system results were compared with the gold standard site of successful ablation or pacing during electrophysiology study and ablation. Mapping time was assessed from time-stamped logs. Prespecified performance goals were used for statistical comparisons.

RESULTS

A total of 255 episodes from 225 patients were enrolled from 4 centers. Regional accuracy for ventricular tachycardia and premature ventricular complexes in patients without significant structural heart disease (n=75, primary end point) was 98.7% (95% CI, 96.0%-100%; P<0.001 to reject predefined H0 <0.80). Regional accuracy for all episodes (secondary end point 1) was 96.9% (95% CI, 94.7%-99.0%; P<0.001 to reject predefined H0 <0.75). Accuracy for the exact or neighboring segment for all episodes (secondary end point 2) was 97.3% (95% CI, 95.2%-99.3%; P<0.001 to reject predefined H0 <0.70). Median spatial accuracy was 15 mm (n=255, interquartile range, 7-25 mm). The mapping process was completed in a median of 0.8 minutes (interquartile range, 0.4-1.4 minutes).

CONCLUSIONS

Computational ECG mapping using a forward-solution approach exceeded prespecified accuracy goals for arrhythmia and pacing localization. Spatial accuracy analysis demonstrated clinically actionable results. This rapid, noninvasive mapping technology may facilitate catheter-based and noninvasive targeted arrhythmia therapies.

REGISTRATION: URL: https://www.

CLINICALTRIALS: gov; Unique identifier: NCT04559061.

Для цитирования: Krummen David E, Villongco Christopher T, Ho Gordon, Schricker Amir A, Field Michael E, Sung Kevin, Kacena Katherine A, Martinson Melissa S, Hoffmayer Kurt S, Hsu Jonathan C, Raissi Farshad, Feld Gregory K, McCulloch Andrew D, Han Frederick T. Forward-Solution Noninvasive Computational Arrhythmia Mapping: The VMAP Study.. Circulation. Arrhythmia and electrophysiology. 2022. DOI: 10.1161/circep.122.010857.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.