Force analysis of bacterial transmission from contact lens cases to corneas, with the contact lens as the intermediary.

Qu Wenwen, Hooymans Johanna M M, de Vries Joop, van der Mei Henny C, Busscher Henk J

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To determine the probability of transmission of a Staphylococcus aureus strain from a contact lens case, to the contact lens (CL) surfaces, to the cornea, on the basis of bacterial adhesion forces measured by using atomic force microscopy (AFM).

METHODS

Adhesion forces between S. aureus strain 835 probes with rigid and soft CLs, storage cases, and porcine corneas were measured with AFM and used to calculate Weibull distributions, from which the transmission probability from one surface to another was derived. Bacterial transmission probabilities from force analyses were compared with experimentally obtained transmission data.

RESULTS

After bond-strengthening, S. aureus adhered to the surface of a lens case with a median force of 10.8 nN. Adhesion forces were different on the soft and rigid CLs (7.7 and 13.6 nN, respectively). Adhesion forces on porcine corneas amounted to 11.8 nN. Data variations were used to calculate the Weibull distribution, from which the probability of transmission from the lens case to a CL and from the CL to the cornea can be directly read. Final transmission probabilities from lens case to the cornea were slightly higher for the rigid (24%) than for the soft (19%) CL. Bacterial transmission determined experimentally increased with increasing contact times, but were within the range of the probabilities derived from Weibull analyses.

CONCLUSIONS

Probabilities of bacterial transmission from contaminated lens cases to corneas can be derived from Weibull analyses of measured forces of adhesion to the surfaces involved.

Для цитирования: Qu Wenwen, Hooymans Johanna M M, de Vries Joop, van der Mei Henny C, Busscher Henk J. Force analysis of bacterial transmission from contact lens cases to corneas, with the contact lens as the intermediary.. Investigative ophthalmology & visual science. 2011. DOI: 10.1167/iovs.10-6392.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.