A European multicientre study on the comparison of HIV-1 viral loads between VERIS HIV-1 Assay and Roche COBAS® TAQMAN® HIV-1 test, Abbott RealTime HIV-1 Assay, and Siemens VERSANT HIV-1 Assay.

Braun Patrick, Delgado Rafael, Drago Monica, Fanti Diana, Fleury Hervé, Hofmann Jörg, Izopet Jacques, Kühn Sebastian, Lombardi Alessandra, Mancon Alessandro, Marcos Mª Angeles, Mileto Davide, Sauné Karine, O'Shea Siobhan, Pérez-Rivilla Alfredo, Ramble John, Trimoulet Pascale, Vila Jordi, Whittaker Duncan, Artus Alain, Rhodes Daniel

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Viral load monitoring is essential for patients under treatment for HIV. Beckman Coulter has developed the VERIS HIV-1 Assay for use on the novel, automated DxN VERIS Molecular Diagnostics System.¥

OBJECTIVES

Evaluation of the clinical performance of the new quantitative VERIS HIV-1 Assay at multiple EU laboratories.

STUDY

DESIGN

Method comparison with the VERIS HIV-1 Assay was performed with 415 specimens at 5 sites tested with COBAS® AmpliPrep/COBAS® TaqMan® HIV-1 Test, v2.0, 169 specimens at 3 sites tested with RealTime HIV-1 Assay, and 202 specimens from 2 sites tested with VERSANT HIV-1 Assay. Patient monitoring sample results from 4 sites were also compared.

RESULTS

Bland-Altman analysis showed the average bias between VERIS HIV-1 Assay and COBAS HIV-1 Test, RealTime HIV-1 Assay, and VERSANT HIV-1 Assay to be 0.28, 0.39, and 0.61 log10 cp/mL, respectively. Bias at low end levels below 1000cp/mL showed predicted bias to be <0.3 log10 cp/mL for VERIS HIV-1 Assay versus COBAS HIV-1 Test and RealTime HIV-1 Assay, and <0.5 log10cp/mL versus VERSANT HIV-1 Assay. Analysis on 174 specimens tested with the 0.175mL volume VERIS HIV-1 Assay and COBAS HIV-1 Test showed average bias of 0.39 log10cp/mL. Patient monitoring results using VERIS HIV-1 Assay demonstrated similar viral load trends over time to all comparators.

CONCLUSIONS

The VERIS HIV-1 Assay for use on the DxN VERIS System demonstrated comparable clinical performance to COBAS® HIV-1 Test, RealTime HIV-1 Assay, and VERSANT HIV-1 Assay.

Для цитирования: Braun Patrick, Delgado Rafael, Drago Monica, Fanti Diana, Fleury Hervé, Hofmann Jörg, Izopet Jacques, Kühn Sebastian, Lombardi Alessandra, Mancon Alessandro, Marcos Mª Angeles, Mileto Davide, Sauné Karine, O'Shea Siobhan, Pérez-Rivilla Alfredo, Ramble John, Trimoulet Pascale, Vila Jordi, Whittaker Duncan, Artus Alain, Rhodes Daniel. A European multicientre study on the comparison of HIV-1 viral loads between VERIS HIV-1 Assay and Roche COBAS® TAQMAN® HIV-1 test, Abbott RealTime HIV-1 Assay, and Siemens VERSANT HIV-1 Assay.. Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology. 2017. DOI: 10.1016/j.jcv.2017.05.003.