DOI 10.1136/neurintsurg-2021-018312

Медицинская библиотека · Journal of neurointerventional surgery · 2022

Evaluation of Sine Spin flat detector CT imaging compared with multidetector CT.

Petroulia Valentina D, Kaesmacher Johannes, Piechowiak Eike I, Dobrocky Tomas, Pilgram-Pastor Sara M, Gralla Jan, Wagner Franca, Mordasini Pasquale

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Flat detector computed tomography (FDCT) is widely used for periprocedural imaging in the angiography suite. Sine Spin FDCT (SFDCT) is the latest generation of cone beam CT using a double oblique trajectory for image acquisition to reduce artefacts and improve soft tissue brain imaging. This study compared the effective dose, image quality and diagnostic performance of the latest generation of SFDCT with multidetector CT (MDCT).

METHODS

An anthropomorphic phantom equipped with MOSFET detectors was used to measure the effective dose of the new 7sDCT Sine Spin protocol on a latest generation biplane angiographic C-arm system. Diagnostic performance was evaluated on periprocedurally acquired SFDCT for depiction of anatomical details, detection of hemorrhage, and ischemia and was compared with preprocedurally acquired MDCT. Inter- and intra-rater correlation as well as sensitivity and specificity were calculated.

RESULTS

Both modalities showed equal diagnostic performance in the supratentorial ventricular system. SFDCT provided inferior image quality in grey-white matter differentiation and infratentorial structures. Intraventricular, subarachnoid and parenchymal hemorrhages were diagnosed with a sensitivity of 83.3%, 84.2% and 75% and a specificity of 97.3%, 80.0% and 100%, respectively; early ischemic lesions with a sensitivity of 73.3% and specificity 94.7%. The effective dose measured for the 7sDCT Sine Spin protocol was 2 mSv.

CONCLUSIONS

Our findings confirm the high diagnostic sensitivity and specificity of SFDCT in detecting intracranial hemorrhage and early ischemic lesions. The delineation of grey-white matter differentiation and infratentorial structures remains a limiting factor. In comparison to previous studies, the new 7sDCT Sine Spin protocol showed a lower effective dose.

Для цитирования: Petroulia Valentina D, Kaesmacher Johannes, Piechowiak Eike I, Dobrocky Tomas, Pilgram-Pastor Sara M, Gralla Jan, Wagner Franca, Mordasini Pasquale. Evaluation of Sine Spin flat detector CT imaging compared with multidetector CT.. Journal of neurointerventional surgery. 2022. DOI: 10.1136/neurintsurg-2021-018312.

Предметные рубрики MeSH