Comparative evaluation of three FDA-approved HIV Ag/Ab combination tests using a genetically diverse HIV panel and diagnostic specimens.

Сравнительное исследованиеОценочное исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Qiu Xiaoxing, Sokoll Lori, Yip Paul, Elliott Debra J, Dua Renu, Mohr Phaedre, Wang Xiao Yan, Spencer Megan, Swanson Priscilla, Dawson George J, Hackett John

Аннотация

BACKGROUND: HIV Ag/Ab combination assays are recommended by CDC for routine screening and several HIV Ag/Ab combination tests are now FDA-approved. Maintaining high specificity and consistent sensitivity across diverse HIV strains is critical for these assays to accurately detect HIV infection and expedite delivery of patient results.

OBJECTIVES: To evaluate performance of three FDA-approved HIV tests: ARCHITECT HIV Combo (Abbott), ADVIA Centaur HIV Combo (Siemens) and BioPlex HIV Ag-Ab (Bio-Rad).

STUDY DESIGN: Sensitivity and specificity were evaluated using an extensive panel of 28 HIV infected human specimens and 17 cultured virus isolates representing multiple genotypes, 6 seroconversion panels, 4 human samples with acute infection, WHO p24 standard and 4020 clinical specimens.

RESULTS: The p24 limit of detection (LOD) for the WHO standard was 0.19IU/ml, 0.70IU/ml, and 1.77IU/ml in BioPlex, ARCHITECT, and Centaur respectively. The distribution of LODs across 15 HIV-1 isolates was substantially narrower in ARCHITECT (5-33pg/ml) than in BioPlex (11-198pg/ml) and Centaur (6-384pg/ml). All assays detected antibodies to the majority of HIV-1 and HIV-2 variants. However, reduced sensitivity was observed for Centaur in detection of antibodies to HIV-1 group M (CRF02_AG), O and N variants. BioPlex and ARCHITECT showed better seroconversion sensitivity than Centaur, detecting one bleed (3-7 days) earlier in 4 (BioPlex) and 3 (ARCHITECT) of 6 seroconversion panels. ARCHITECT demonstrated the highest specificity (99.90-100%) compared to BioPlex (99.80%) and Centaur (99.42%).

CONCLUSIONS: The overall performance of ARCHITECT and BioPlex was superior to Centaur, especially for detection of acute HIV infection.

Предметные рубрики MeSH