His-bundle pacing vs biventricular pacing following atrioventricular nodal ablation in patients with atrial fibrillation and reduced ejection fraction: A multicenter, randomized, crossover study-The ALTERNATIVE-AF trial.

Huang Weijian, Wang Songjie, Su Lan, Fu Guosheng, Su Yangang, Chen Keping, Zou Jiangang, Han Hongwei, Wu Shengjie, Sheng Xia, Chen Xueying, Fan Xiaohan, Xu Lei, Zhou Xiaohong, Mao Guangyun, Ellenbogen Kenneth A, Whinnett Zachary I

Рандомизированное исследованиеМногоцентровое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Atrioventricular nodal ablation (AVNA) combined with biventricular pacing (BVP) improves outcomes in patients with persistent atrial fibrillation (AF), adequate rate control, and reduced left ventricular ejection fraction (LVEF). His-bundle pacing (HBP) delivers physiological ventricular activation and is a promising alternative to BVP.

OBJECTIVE

The purpose of this trial was to compare HBP with BVP following AVNA.

METHODS

In this multicenter, prospective, randomized crossover trial, we recruited patients with persistent AF and reduced LVEF (≤40%). All patients underwent AVNA and received both HBP and BVP. Patients were randomized to either HBP or BVP for 9 months (phase 1), then were switched to the alternative pacing modality for the next 9 months (phase 2). The primary endpoint was change in LVEF.

RESULTS

Fifty patients (age 64.3 ± 10.3 years; ventricular rate 93.1 ± 19.9 bpm; 72% male) were enrolled. Thirty-eight patients completed the 2 phases and were included in the crossover analysis. A significant improvement in LVEF was observed with HBP compared to BVP (phase 1: ΔLVEFHBP 21.3% and ΔLVEFBVP 16.7%; phase 2: ΔLVEFHBP 3.5% and ΔLVEFBVP -2.4%; Pgeneralizedadditivemodel = 0.015). Significant improvements in left ventricular end-diastolic diameter, New York Heart Association functional class, and B-type natriuretic peptide level were observed with both pacing modalities compared with baseline, whereas no significant differences were observed between HBP and BVP.

CONCLUSION

HBP delivers a modest but significant improvement in LVEF in patients with persistent AF, impaired ventricular function, and narrow QRS duration post-AVNA compared with BVP. Larger long-term trials are required to confirm the additional improvements in function with HBP.

Для цитирования: Huang Weijian, Wang Songjie, Su Lan, Fu Guosheng, Su Yangang, Chen Keping, Zou Jiangang, Han Hongwei, Wu Shengjie, Sheng Xia, Chen Xueying, Fan Xiaohan, Xu Lei, Zhou Xiaohong, Mao Guangyun, Ellenbogen Kenneth A, Whinnett Zachary I. His-bundle pacing vs biventricular pacing following atrioventricular nodal ablation in patients with atrial fibrillation and reduced ejection fraction: A multicenter, randomized, crossover study-The ALTERNATIVE-AF trial.. Heart rhythm. 2022. DOI: 10.1016/j.hrthm.2022.07.009.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.