Investigating the effect of thyroid shields and lead glasses on image quality and artefacts in cone beam computed tomography devices: A phantom study.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Hafezi L, Aghaz A, Deevband M R, Vaezi S
Аннотация
INTRODUCTION: The objective of this study is to investigate and compare the impact of thyroid shields and lead glasses on the image quality (IQ) and artefacts in dental Cone Beam CT (CBCT) images. The current study aims to evaluate the effectiveness of this protective equipment in enhancing the quality of dental CBCT images.
METHODS: In this study, a New Tom dental CBCT device was used to scan Pro-Project phantom with and without thyroid shields and lead glasses in three different large, medium, and small Fields of View (FOVs). The several key IQ parameters such as noise, contrast-to-noise ratio (CNR), homogeneity, contrast-resolution (CR), and spatial-resolution (SR), were evaluated using ImageJ software. The measurements were performed and compared in various cross-sections based on the location of test objects in the phantom.
RESULTS: The study found that the use of thyroid shields and lead glasses resulted in a decrease in CNR and homogeneity parameters, while noise increased. However, there were no significant changes in SR and CR. It should be noted that the use of thyroid shields and lead glasses may increase image noise by 1-7 %. The combined use of lead shields was not included in our findings, as their effects are localized and similar to separate evaluations in adjacent areas in lower and upper sections.
CONCLUSION: Thyroid shields and lead glasses during dental CBCT imaging increase image noise and reduce CNR. Proper placement is crucial, and reducing the FOV size can minimise artefacts while preserving diagnostic accuracy.
IMPLICATIONS FOR PRACTICE: Thyroid shields and lead glasses are vital for radiation protection. Increased noise demands attention in clinical practice. Ongoing research, training, quality assurance, and effective patient communication are essential for safety and trust in dental imaging.
METHODS: In this study, a New Tom dental CBCT device was used to scan Pro-Project phantom with and without thyroid shields and lead glasses in three different large, medium, and small Fields of View (FOVs). The several key IQ parameters such as noise, contrast-to-noise ratio (CNR), homogeneity, contrast-resolution (CR), and spatial-resolution (SR), were evaluated using ImageJ software. The measurements were performed and compared in various cross-sections based on the location of test objects in the phantom.
RESULTS: The study found that the use of thyroid shields and lead glasses resulted in a decrease in CNR and homogeneity parameters, while noise increased. However, there were no significant changes in SR and CR. It should be noted that the use of thyroid shields and lead glasses may increase image noise by 1-7 %. The combined use of lead shields was not included in our findings, as their effects are localized and similar to separate evaluations in adjacent areas in lower and upper sections.
CONCLUSION: Thyroid shields and lead glasses during dental CBCT imaging increase image noise and reduce CNR. Proper placement is crucial, and reducing the FOV size can minimise artefacts while preserving diagnostic accuracy.
IMPLICATIONS FOR PRACTICE: Thyroid shields and lead glasses are vital for radiation protection. Increased noise demands attention in clinical practice. Ongoing research, training, quality assurance, and effective patient communication are essential for safety and trust in dental imaging.