MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/j.1540-8167.1998.tb00141.x

Медицинская библиотека · Journal of cardiovascular electrophysiology · 1998

Optimization of shocking lead configuration for transvenous atrial defibrillation.

Krum D, Hare J, Mughal K, Jazayeri M R, Deshpande S, Dhala A, Blanck Z, Akhtar M, Sra J

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

High atrial defibrillation energy requirements (ADER) in patients with chronic atrial fibrillation (AF) may limit the acceptance of transvenous atrial defibrillation. We evaluated an optimized defibrillation electrode configuration that could help to reduce the ADER in patients with AF.

METHODS AND

RESULTS

We tested ten different configurations in nine dogs with AF (3.33+/-2.92 days) induced by rapid atrial pacing. The configurations were: right atrial (RA) appendage as anode and coronary sinus (CS) as cathode; RA and innominate vein (I) as anode to CS (cathode); RA-CS (anode) to I (cathode); I-CS (anode) to RA (cathode); RA and left lateral subcutaneous patch (P) as anode to CS (cathode); RA-CS (anode) to P (cathode); P-CS (anode) to RA (cathode); superior vena cava (SVC) and CS (anode) to RA (cathode); RA-CS (anode) to SVC (cathode); and RA-SVC (anode) to CS (cathode). ADER was defined as the voltage needed to defibrillate the atria in 10% to 90% of 20 consecutive shocks. Three lead systems had ADER lower than the RA (anode) to CS (cathode) configuration, which required a mean of 143+/-58 volts. These three were: RA-SVC (anode) to CS (cathode) 103+/-29 V; I-CS (anode) to RA (cathode) 129+/-39 V; and P-CS (anode) to RA (cathode) 130+/-38 V. The remaining configurations had ADER higher than the RA (anode) to CS (cathode) configuration.

CONCLUSION

Adding an additional shocking electrode may reduce ADER when compared with the RA (anode) to CS (cathode) configuration. This concept could be incorporated into future implantable atrial defibrillators or used for refractory patients undergoing temporary transvenous cardioversion.

Для цитирования: Krum D, Hare J, Mughal K, Jazayeri M R, Deshpande S, Dhala A, Blanck Z, Akhtar M, Sra J. Optimization of shocking lead configuration for transvenous atrial defibrillation.. Journal of cardiovascular electrophysiology. 1998. DOI: 10.1111/j.1540-8167.1998.tb00141.x.

Предметные рубрики MeSH