DOI 10.1007/s10029-025-03580-9
Concurrent validity and agreement of the HerniaCare Lab device for abdominal wall strength assessment.
Аннотация
PURPOSE: Currently, no standardized, low-cost, and portable method is available for assessing abdominal wall strength in patients with incisional hernias, addressing the limitations of traditional dynamometry.
METHODS
This cross-sectional validation study compared the HerniaCare Lab device performance with the Activforce 2 hand-held dynamometer in92 adults diagnosed with abdominal wall hernias. Isometric trunk flexion strength was measured under identical conditions, and agreement between devices was analyzed using non-parametric tests, correlation, and concordance statistics.
RESULTS
The HerniaCare Lab showed systematically higher strength values than the Activforce 2 (mean difference = + 22.3 N, p < 0.001) but demonstrated a very strong positive correlation (ρ = 0.95, p < 0.001) and good concordance (CCC = 0.89). Bland–Altman analysis revealed a mean bias of + 24.9 N with 95% limits of agreement from − 17.7 to+ 67.5 N, and a slight proportional bias at higher force levels.
CONCLUSION
Despite predictable overestimation, the HerniaCare Lab exhibits strong concurrent validity and good agreement with an established reference device, supporting its potential clinical utility for objective, reproducible, and accessible assessment of abdominal wall strength in surgical populations.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.