Mitigation of Ventilator-induced Diaphragm Atrophy by Transvenous Phrenic Nerve Stimulation.

Reynolds Steven C, Meyyappan Ramasamy, Thakkar Viral, Tran Bao D, Nolette Marc-André, Sadarangani Gautam, Sandoval Rodrigo A, Bruulsema Laura, Hannigan Brett, Li Jason W, Rohrs Elizabeth, Zurba Jason, Hoffer Joaquín Andrés

СтатьяПеревод пока не подготовлен

Аннотация

RATIONALE: Ventilator-induced diaphragm dysfunction is a significant contributor to weaning difficulty in ventilated critically ill patients. It has been hypothesized that electrically pacing the diaphragm during mechanical ventilation could reduce diaphragm dysfunction.

OBJECTIVES

We tested a novel, central line catheter-based, transvenous phrenic nerve pacing therapy for protecting the diaphragm in sedated and ventilated pigs.

METHODS

Eighteen Yorkshire pigs were studied. Six pigs were sedated and mechanically ventilated for 2.5 days with pacing on alternate breaths at intensities that reduced the ventilator pressure-time product by 20-30%. Six matched subjects were similarly sedated and ventilated but were not paced. Six pigs served as never-ventilated, never-paced control animals.

MEASUREMENTS AND MAIN

RESULTS

Cumulative duration of pacing therapy ranged from 19.7 to 35.7 hours. Diaphragm thickness assessed by ultrasound and normalized to initial value showed a significant decline in ventilated-not paced but not in ventilated-paced subjects (0.84 [interquartile range (IQR), 0.78-0.89] vs. 1.10 [IQR, 1.02-1.24]; P = 0.001). Compared with control animals (24.6 μm2/kg; IQR, 21.6-26.0), median myofiber cross-sectional areas normalized to weight and sarcomere length were significantly smaller in the ventilated-not paced (17.9 μm2/kg; IQR, 15.3-23.7; P = 0.005) but not in the ventilated-paced group (24.9 μm2/kg; IQR, 16.6-27.3; P = 0.351). After 60 hours of mechanical ventilation all six ventilated-paced subjects tolerated 8 minutes of intense phrenic stimulation, whereas three of six ventilated-not paced subjects did not (P = 0.055). There was a nonsignificant decrease in diaphragm tetanic force production over the experiment in the ventilated-paced and ventilated-not paced groups.

CONCLUSIONS

These results suggest that early transvenous phrenic nerve pacing may mitigate ventilator-induced diaphragm dysfunction.

Для цитирования: Reynolds Steven C, Meyyappan Ramasamy, Thakkar Viral, Tran Bao D, Nolette Marc-André, Sadarangani Gautam, Sandoval Rodrigo A, Bruulsema Laura, Hannigan Brett, Li Jason W, Rohrs Elizabeth, Zurba Jason, Hoffer Joaquín Andrés. Mitigation of Ventilator-induced Diaphragm Atrophy by Transvenous Phrenic Nerve Stimulation.. American journal of respiratory and critical care medicine. 2017. DOI: 10.1164/rccm.201502-0363oc.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.