Evaluation of respiratory rate monitoring performance using a home oxygen monitoring device among patients with interstitial lung disease and chronic obstructive pulmonary disease.

Hamada Satoshi, Handa Tomohiro, Tanabe Naoya, Sato Susumu, Tanizawa Kiminobu, Sato Atsuyasu, Morita Satoshi, Chin Kazuo, Hirai Toyohiro

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Home monitoring devices have been developed to measure adherence to home oxygen therapy. In this study, we evaluated the performance of TeleOx®, a commercially available remote monitoring device, in comparison with polysomnography (PSG) in patients with interstitial lung disease (ILD) and chronic obstructive pulmonary disease (COPD) and the factors that affected TeleOx® correct use.

METHODS

TeleOx® was connected on the patient or concentrator side. The oxygen flow rates were set at 1, 3, and 5 L/min. Intraclass correlation coefficient (ICC) (2,1) was used to determine the agreement between respiratory rate measured using TeleOx® and that measured using PSG, and the minimum acceptable level of reliability was >0.7.

RESULTS

In total, 22 patients (16 with ILD and 6 with COPD) were assessed. In patients with ILD, the detection rate of patients' respiration assessed using TeleOx® did not change according to the device's position. It increased from 53.5% to 79.0% by changing the position from the concentrator to the patient side in patients with COPD. The ICC (2,1) value indicated that TeleOx® had acceptable reliability at oxygen flow rates of 1 and 3 L/min regardless of the device's position in patients with ILD (the concentrator side: 0.9 and 0.82, respectively; the patient side: 0.95 and 0.82, respectively), whereas that did only at the oxygen flow rate of 1 L/min and in connecting TeleOx® on the patient side in patients with COPD (0.73).

CONCLUSION

The monitoring performance of TeleOx® differed according to its position, oxygen flow rates, and patients' diseases.

Для цитирования: Hamada Satoshi, Handa Tomohiro, Tanabe Naoya, Sato Susumu, Tanizawa Kiminobu, Sato Atsuyasu, Morita Satoshi, Chin Kazuo, Hirai Toyohiro. Evaluation of respiratory rate monitoring performance using a home oxygen monitoring device among patients with interstitial lung disease and chronic obstructive pulmonary disease.. Sarcoidosis, vasculitis, and diffuse lung diseases : official journal of WASOG. 2022. DOI: 10.36141/svdld.v39i1.12133.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.