Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2018

Impact of simulated MitraClip on forward flow obstruction in the setting of mitral leaflet tethering: An in vitro investigation.

Сравнительное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Bloodworth Charles H, Pierce Eric L, Kohli Keshav, Deaton Nancy J, Jones Kaitlin J, Duvvuri Radhika, Kamioka Norihiko, Babaliaros Vasilis C, Yoganathan Ajit P

Аннотация

OBJECTIVES: We aimed to evaluate diastolic leaflet tethering as a factor that may cause mitral stenosis (MS) after simulated MitraClip implantation, using an in vitro left heart simulator.

BACKGROUND: Leaflet tethering commonly seen in functional mitral regurgitation may be a significant factor affecting the severity of MS after MitraClip implantation.

METHODS: A left heart simulator with excised ovine mitral valves (N = 6), and custom edge-to-edge clip devices (GTclip) was used to mimic implantation of MitraClip in a variety of positions. Anterior mitral leaflet (AML) tethering severity was varied for each case (leaflet excursion of 75°, 60°, and 45°, consistent with mild, moderate and severe tethering), and the baseline mitral annular area (MAA) was varied across samples (3.6-4.8 cm2 ). The resulting mitral valve area (MVA), and peak/mean mitral valve gradient (MVG) were measured in each case.

RESULTS: AML tethering severity was a highly significant factor increasing MVG and decreasing MVA (P < 0.001). When GTclip placement was simulated with severe AML tethering, mean MVG >5 mmHg resulted more frequently than with GTclip placement alone (46% vs. 4%, respectively). However, severe AML tethering alone significantly reduced baseline MVA to 3.6 ± 0.2 cm2 , and increased baseline MVG to 3.0 ± 0.4 mmHg. At MAA above 4.7 cm2 , severe AML tethering did not cause moderate MS, even with placement of two GTclips (95% confidence).

CONCLUSIONS: Our results show that diastolic AML tethering may predispose to MS after clip placement, however, MS was not observed when baseline MVA was above 4.0 cm2 . Severity of AML tethering may be an important criterion in selecting patients for edge-to-edge repair.