Reuse of insulin pens among multiple patients at 2 Veterans Affairs medical centers.

Schirmer Patricia, Winston Carla A, Lucero-Obusan Cynthia, Winters Mark, Lesse Alan, de Comarmond Charles, Oda Gina, Martinello Richard A, Holodniy Mark

СтатьяПоддержка правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To determine whether reuse of insulin pens among multiple patients resulted in transmission of bloodborne pathogens (BBP).

DESIGN

Retrospective cohort study.

SETTING

Two Veterans Affairs medical centers.

PATIENTS: Veterans who received insulin via insulin pens from 2010 to 2013. METHODS Patients were identified through electronic health records, notified of possible exposure, and serotested for human immunodeficiency virus, hepatitis C virus (HCV), and hepatitis B virus. Newly discovered case patients were assessed in relation to potential proximate patients to determine viral strain relatedness by HCV envelope (env) gene sequencing.

RESULTS

Of 1,791 hospitalized veterans who received insulin via insulin pen, 1,155 were tested for at least 1 viral infection after exposure. Of these, 67 patients were newly diagnosed with 1 or more viral BBPs. For human immunodeficiency virus and hepatitis B virus no additional strain testing of case or proximate patients was possible; 8 HCV cases and 45 proximates (40 unique patients; 5 patients were positive for 2 genotypes) were identified as needing strain testing. Only 3 cases and their 19 proximates had samples available for further testing. None of the 26 remaining proximate patients had blood available for further testing. Median genetic distance between the HCV env sequences of those available for additional testing ranged from 14% to 24%, indicating nonrelatedness.

CONCLUSIONS

Our investigation revealed that exposure to insulin pen reuse did not result in HCV transmission among patients who had viral genetic analysis performed. Analysis for any additional potential transmission of blood-borne pathogens was limited by the available samples.

Для цитирования: Schirmer Patricia, Winston Carla A, Lucero-Obusan Cynthia, Winters Mark, Lesse Alan, de Comarmond Charles, Oda Gina, Martinello Richard A, Holodniy Mark. Reuse of insulin pens among multiple patients at 2 Veterans Affairs medical centers.. Infection control and hospital epidemiology. 2015. DOI: 10.1017/ice.2015.165.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.