DOI 10.2337/dc22-0073

Медицинская библиотека · Diabetes care · 2022

Accuracy of a Factory-Calibrated Continuous Glucose Monitor in Individuals With Diabetes on Hemodialysis.

Villard Orianne, Breton Marc D, Rao Swati, Voelmle Mary K, Fuller Morgan R, Myers Helen E, McFadden Ryan K, Luke Zander S, Wakeman Christian A, Clancy-Oliveri Mary, Basu Ananda, Stumpf Meaghan M

СтатьяПоддержка исследования: вне правительства СШАОбзорПеревод пока не подготовлен

Аннотация

OBJECTIVE

Continuous glucose monitoring (CGM) improves diabetes management, but its reliability in individuals on hemodialysis is poorly understood and potentially affected by interstitial and intravascular volume variations.

RESEARCH DESIGN AND

METHODS

We assessed the accuracy of a factory-calibrated CGM by using venous blood glucose measurements (vBGM) during hemodialysis sessions and self-monitoring blood glucose (SMBG) at home.

RESULTS

Twenty participants completed the protocol. The mean absolute relative difference of the CGM was 13.8% and 14.4%, when calculated on SMBG (n = 684) and on vBGM (n = 624), and 98.7% and 100% of values in the Parkes error grid A/B zones, respectively. Throughout 181 days of CGM monitoring, the median time in range (70-180 mg/dL) was 38.5% (interquartile range 29.3-57.9), with 28.7% (7.8-40.6) of the time >250 mg/dL.

CONCLUSIONS

The overall performance of a factory-calibrated CGM appears reasonably accurate and clinically relevant for use in practice by individuals on hemodialysis and health professionals to improve diabetes management.

Для цитирования: Villard Orianne, Breton Marc D, Rao Swati, Voelmle Mary K, Fuller Morgan R, Myers Helen E, McFadden Ryan K, Luke Zander S, Wakeman Christian A, Clancy-Oliveri Mary, Basu Ananda, Stumpf Meaghan M. Accuracy of a Factory-Calibrated Continuous Glucose Monitor in Individuals With Diabetes on Hemodialysis.. Diabetes care. 2022. DOI: 10.2337/dc22-0073.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.