MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/03000605221108100

Медицинская библиотека · The Journal of international medical research · 2022

Correlation between corneal dynamic responses and keratoconus topographic parameters.

Tai Hsi-Yun, Lin Jun-Ji, Huang Yi-Hung, Shih Po-Jen, Wang I-Jong, Yen Jia-Yush

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To investigate the correlation between corneal biomechanical properties and topographic parameters using machine learning networks for automatic severity diagnosis and reference benchmark construction.

METHODS

This was a retrospective study involving 31 eyes from 31 patients with keratonus. Two clustering approaches were used (i.e., shape-based and feature-based). The shape-based method used a keratoconus benchmark validated for indicating the severity of keratoconus. The feature-based method extracted imperative features for clustering analysis.

RESULTS

There were strong correlations between the symmetric modes and the keratoconus severity and between the asymmetric modes and the location of the weak centroid. The Pearson product-moment correlation coefficient (PPMC) between the symmetric mode and normality was 0.92 and between the asymmetric mode and the weak centroid value was 0.75.

CONCLUSION

This study confirmed that there is a relationship between the keratoconus signs obtained from topography and the corneal dynamic behaviour captured by the Corvis ST device. Further studies are required to gather more patient data to establish a more extensive database for validation.

Для цитирования: Tai Hsi-Yun, Lin Jun-Ji, Huang Yi-Hung, Shih Po-Jen, Wang I-Jong, Yen Jia-Yush. Correlation between corneal dynamic responses and keratoconus topographic parameters.. The Journal of international medical research. 2022. DOI: 10.1177/03000605221108100.