DOI 10.1097/j.jcrs.0000000000000977

Медицинская библиотека · Journal of cataract and refractive surgery · 2022

Measuring corneal astigmatism using OCT in keratoconus.

Llorens-Quintana Clara, Lee Daniel J, Pavlatos Elias, Chamberlain Winston, Huang David, Li Yan

Наблюдательное исследованиеСтатьяПоддержка исследования: вне правительства СШАПоддержка NIH: внешняя программаПеревод пока не подготовлен

Аннотация

PURPOSE: To measure net corneal astigmatism using optical coherence tomography (OCT) (Avanti) in individuals with keratoconus and compare the repeatability and accuracy with those obtained using Scheimpflug imaging (Pentacam HR).

SETTING

Casey Eye Institute, Portland, Oregon.

DESIGN

Prospective cross-sectional observational study.

METHODS

Net corneal astigmatism was calculated in keratoconic and normal eyes using OCT and Scheimpflug imaging with 4 settings-pupil or vertex centration settings with a 3 or 4 mm circular analytical zone. Corneal elevation maps were obtained from OCT images and fitted with the Zernike polynomials to obtain net corneal astigmatism. Manifest refraction astigmatism was used to evaluate the accuracy of net corneal astigmatism measurements. The coefficient of repeatability from 2 repeated measures was calculated.

RESULTS

46 eyes with manifest or subclinical keratoconus and 52 normal control eyes were analyzed. For OCT measurements in keratoconus, better accuracy was achieved with pupil centration and 3 mm analytical zone; however, better repeatability was achieved with vertex centration and 4 mm analytical zone (coefficient of repeatability = 0.53 diopters, the Fligner-Killeen test with Bonferroni adjustment P < .0017). Agreement with manifest refraction was significantly better with OCT compared with that using Pentacam HR (generalized mixed-effect model with Bonferroni adjustment P < .00625). No statistically significant difference was found between instruments or settings in control eyes.

CONCLUSIONS

OCT was able to measure net corneal astigmatism with better accuracy and precision in keratoconic eyes than the Pentacam HR. Measurements may be more accurate using pupil centration and a smaller analytical zone in patients with keratoconus.

Для цитирования: Llorens-Quintana Clara, Lee Daniel J, Pavlatos Elias, Chamberlain Winston, Huang David, Li Yan. Measuring corneal astigmatism using OCT in keratoconus.. Journal of cataract and refractive surgery. 2022. DOI: 10.1097/j.jcrs.0000000000000977.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.