DOI 10.1371/journal.pone.0268856

Медицинская библиотека · PloS one · 2022

Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients.

Asawaworarit Rosepon, Satitpitakul Vannarut, Taweekitikul Parichart, Pongpirul Krit

СтатьяНаблюдательное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate agreement of total corneal power (TCP) measured by swept-source anterior segment optical coherence tomography (CASIA 2), Scheimpflug tomography (Pentacam AXL), and swept-source optical biometer (IOLMaster 700) in normal and keratoconic patients.

METHODS

This is a prospective observational study conducted at King Chulalongkorn Memorial Hospital, Bangkok, Thailand. Biometric values were measured by each device for three times by two operators to evaluate repeatability and reproducibility of TCP. The agreement of TCP and other parameters including total corneal astigmatism, anterior keratometry, anterior corneal astigmatism, posterior keratometry, posterior corneal astigmatism, anterior chamber depth, white-to-white corneal diameter (WTW), central corneal thickness, and intraocular power were also evaluated.

RESULTS

This study enrolled 100 healthy participants and 34 patients with keratoconus. The repeatability and reproducibility of TCP were high in all devices (ICC > 0.9). The agreement of TCP was excellent among three devices in both groups (ICC > 0.9). However, the agreement of TCP between CASIA 2 and IOLMaster 700 was slightly lower in healthy participants (ICC = 0.867) and keratoconic patients (ICC = 0.852) with mean differences of more than 1.0 diopter is clinically significant. Wider 95% limit of agreement were found in keratoconic eyes. Most of other parameters showed good to excellent agreement except WTW which showed poor to moderate agreement in both groups. IOL power showed clinically significant differences in patients with keratoconus.

CONCLUSIONS

The agreement of TCP measured by three devices was excellent in normal and keratoconic patients. However, TCP cannot be used interchangeably between devices.

Для цитирования: Asawaworarit Rosepon, Satitpitakul Vannarut, Taweekitikul Parichart, Pongpirul Krit. Agreement of total corneal power between 2 swept-source optical coherence tomography and Scheimpflug tomography in normal and keratoconic patients.. PloS one. 2022. DOI: 10.1371/journal.pone.0268856.