Cerebral intolerance during flow arrested carotid angioplasty.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
St Louis Myron, Park Brian D, Dahn Michael, Bozeman Patricia
Аннотация
INTRODUCTION: The use of flow arrest as a means of providing cerebral protection during carotid angioplasty offers the advantages of improved efficiency of debris removal and the ability to provide protection under unfavorable (tortuous) anatomic circumstances. However, in contrast to the filtration methods of cerebral protection, this modality requires complete interruption of antegrade carotid artery flow during balloon angioplasty and stent deployment.
METHODS: We report our experience with 9 patients undergoing carotid angioplasty with the Mo.Ma device, which utilizes common and external carotid artery balloon occlusion during the angioplasty procedure. We assessed the clinical outcomes and intraprocedural hemodynamic data.
RESULTS: The average duration of carotid occlusion was 8.3 minutes. Of the 9 patients, 2 patients (22%) experienced cerebral intolerance. No stroke occurred in this patient cohort. There appeared to be a poor relationship between procedure intolerance and the presence of significant contralateral stenosis or low carotid back pressure. Furthermore, the incidence of postangioplasty hypotension was not clearly related to cerebral intolerance.
CONCLUSION: Carotid angioplasty with stenting can be safely conducted with flow arrest as an alternative to filter-type cerebral protection devices. However, because cerebral intolerance is not an infrequent occurrence with this approach, clinicians must be cognizant of management strategies for transient cerebral intolerance.
METHODS: We report our experience with 9 patients undergoing carotid angioplasty with the Mo.Ma device, which utilizes common and external carotid artery balloon occlusion during the angioplasty procedure. We assessed the clinical outcomes and intraprocedural hemodynamic data.
RESULTS: The average duration of carotid occlusion was 8.3 minutes. Of the 9 patients, 2 patients (22%) experienced cerebral intolerance. No stroke occurred in this patient cohort. There appeared to be a poor relationship between procedure intolerance and the presence of significant contralateral stenosis or low carotid back pressure. Furthermore, the incidence of postangioplasty hypotension was not clearly related to cerebral intolerance.
CONCLUSION: Carotid angioplasty with stenting can be safely conducted with flow arrest as an alternative to filter-type cerebral protection devices. However, because cerebral intolerance is not an infrequent occurrence with this approach, clinicians must be cognizant of management strategies for transient cerebral intolerance.