A comparison between magnetic resonance angiography at 3 Teslas (time-of-flight and contrast-enhanced) and flat-panel digital subtraction angiography in the assessment of embolized brain aneurysms.

Nakiri Guilherme S, Santos Antonio C, Abud Thiago G, Aragon Davi C, Colli Benedicto O, Abud Daniel G

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the time-of-flight and contrast-enhanced- magnetic resonance angiography techniques in a 3 Tesla magnetic resonance unit with digital subtraction angiography with the latest flat-panel technology and 3D reconstruction in the evaluation of embolized cerebral aneurysms.

INTRODUCTION

Many embolized aneurysms are subject to a recurrence of intra-aneurismal filling. Traditionally, imaging surveillance of coiled aneurysms has consisted of repeated digital subtraction angiography. However, this method has a small but significant risk of neurological complications, and many authors have advocated the use of noninvasive imaging methods for the surveillance of embolized aneurysms.

METHODS

Forty-three aneurysms in 30 patients were studied consecutively between November 2009 and May 2010. Two interventional neuroradiologists rated the time-of-flight-magnetic resonance angiography, the contrast-enhanced-magnetic resonance angiography, and finally the digital subtraction angiography, first independently and then in consensus. The status of aneurysm occlusion was assessed according to the Raymond scale, which indicates the level of recanalization according to degrees: Class 1: excluded aneurysm; Class 2: persistence of a residual neck; Class 3: persistence of a residual aneurysm. The agreement among the analyses was assessed by applying the Kappa statistic.

RESULTS

Inter-observer agreement was excellent for both methods (K = 0.93; 95 % CI: 0.84-1). Inter-technical agreement was almost perfect between time-of-flight-magnetic resonance angiography and digital subtraction angiography (K = 0.98; 95 % CI: 0.93-1) and between time-of-flight-magnetic resonance angiography and contrast-enhanced-magnetic resonance angiography (K = 0.98; 95% CI: 0.93-1). Disagreement occurred in only one case (2.3%), which was classified as Class I by time-of-flight-magnetic resonance angiography and Class II by digital subtraction angiography. The agreement between contrast-enhanced-magnetic resonance angiography and digital subtraction angiography was perfect (K = 1; 95% CI: 1-1). In three patients, in-stent stenosis was identified by magnetic resonance angiography but not confirmed by digital subtraction angiography.

CONCLUSION

Digital subtraction angiography and both 3T magnetic resonance angiography techniques have excellent reproducibility for the assessment of aneurysms embolized exclusively with coils. In those cases also treated with stent remodeling, digital subtraction angiography may still be necessary to confirm eventual parent artery stenosis, as identified by magnetic resonance angiography.

Для цитирования: Nakiri Guilherme S, Santos Antonio C, Abud Thiago G, Aragon Davi C, Colli Benedicto O, Abud Daniel G. A comparison between magnetic resonance angiography at 3 Teslas (time-of-flight and contrast-enhanced) and flat-panel digital subtraction angiography in the assessment of embolized brain aneurysms.. Clinics (Sao Paulo, Brazil). 2011. DOI: 10.1590/s1807-59322011000400020.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.