DOI 10.1002/jcla.24029

Медицинская библиотека · Journal of clinical laboratory analysis · 2021

Developing an ultra-performance liquid chromatography-tandem mass spectrometry for detecting aldosterone in human plasma.

Lin Wenbin, Yao Zhenrong, Li Yuzhe, Liao Zhihao, Xiao Jiahao, Chen Yonghong, Feng Pinning, Gan Wenjia

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Accurately measuring plasma aldosterone concentration is difficult but meaningful for primary aldosteronism (PA) diagnosis.

METHODS

In this study, we developed an ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) method for plasma aldosterone detection, evaluated its performance according to guidelines issued by CLSI, including detection limit, linearity, precision, and compared it with chemiluminescence immunoassay. Then, a reference range of plasma aldosterone in young people was established by using this method.

RESULTS

The lower limit of quantitation (LOQ) was 10 pg/ml. The mean recovery rates of analyte added to serum were 100.07-102.05% in different concentrations. The linearity range was 20-2000 pg/ml. Inter-assay CVs were 2.20-3.97% at aldosterone concentrations of 65.66-854.75 pg/ml. The regression equation of UPLC-MS/MS (x) and chemiluminescence immunoassay (y) was y = 1.002x + 65.854 (r = 0.9456, n = 237). The reference range of plasma aldosterone detected by UPLC-MS/MS was 11.30-363.82 pg/ml in young people in South China, and there was no statistically significant difference in plasma aldosterone concentration between two genders.

CONCLUSION

In conclusion, UPLC-MS/MS can rapidly and accurately detect plasma aldosterone and is appropriate for clinical application.

Для цитирования: Lin Wenbin, Yao Zhenrong, Li Yuzhe, Liao Zhihao, Xiao Jiahao, Chen Yonghong, Feng Pinning, Gan Wenjia. Developing an ultra-performance liquid chromatography-tandem mass spectrometry for detecting aldosterone in human plasma.. Journal of clinical laboratory analysis. 2021. DOI: 10.1002/jcla.24029.