Enhanced detection and Characterization of M-proteins in multiple myeloma patients using An Agilent AssayMAP Bravo liquid handling system coupled to an LC-QTOF.

Nichols Matthew, Phua Chai W, Louzada Martha L, Hedley Benjamin D, Bhayana Vipin, Chin-Yee Ian, Rutledge Angela C

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Mass spectrometry methods are emerging as tools to detect M-proteins in the serum of multiple myeloma patients with increased sensitivity and specificity compared to traditional electrophoretic methods.

METHODS

A liquid handling system, the Agilent AssayMAP Bravo, with liquid chromatography high-resolution quadrupole-time-of-flight (LC-QTOF) mass spectrometry to analyze intact light chains was compared to immunofixation electrophoresis (IFE) for M-protein analysis. 210 patient serum samples were analyzed in a split sample comparison (LC-QTOF vs. IFE). LC-QTOF and IFE were interpreted by different individuals in a blinded fashion and results were grouped into four categories: IFE+/QTOF+, IFE+/QTOF-, IFE-/QTOF+, or IFE-/QTOF-.

RESULTS

The LC-QTOF method is able to determine the isotype of M-proteins in a similar fashion to IFE. The estimated limit of detection was ∼ 35 mg/L for adalimumab. For split patient samples, 168 were QTOF+/IFE+, 25 were QTOF-/IFE-, 14 were QTOF+/IFE-, and three were QTOF-/IFE + . Excluding the QTOF+/IFE- results due to the improved sensitivity of the LC-QTOF method, the concordance of LC-QTOF with IFE was ∼ 98 %. The LC-QTOF method also offers improved specificity compared to electrophoretic methods due to inclusion of the accurate mass of the light chains.

CONCLUSIONS

The LC-QTOF method was deemed fit for clinical use as a qualitative test with increased sensitivity and specificity compared to IFE. The LC-QTOF can also better resolve therapeutic and multiple myeloma IgG-kappa M-proteins, which present a challenge for electrophoretic methods. Future work will determine suitability of this method as an assessment of minimal residual disease status.

Для цитирования: Nichols Matthew, Phua Chai W, Louzada Martha L, Hedley Benjamin D, Bhayana Vipin, Chin-Yee Ian, Rutledge Angela C. Enhanced detection and Characterization of M-proteins in multiple myeloma patients using An Agilent AssayMAP Bravo liquid handling system coupled to an LC-QTOF.. Clinical biochemistry. 2024. DOI: 10.1016/j.clinbiochem.2024.110870.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.