Paraprotein Interferences: Insights from a Short Study Involving Multiple Platforms and Multiple Measurands.
Аннотация
BACKGROUND
Though paraproteinaemic interferences is a well-known phenomenon in clinical chemistry, a large-scale evaluation study involving multiple paraproteinaemic specimens on multiple platforms including multiple measurands with an aim to provide a predictive analysis, is singularly lacking. The present study aims to fill this gap in research.
MATERIAL AND
METHODS
This cross-sectional non-interventional observational study involved thirteen paraproteinaemic subjects, determined their gamma globulin characterization and measured their total bilirubin, direct bilirubin, HDL-cholesterol, calcium, inorganic phosphate, iron and unsaturated iron binding capacity (UIBC) levels on a dry chemistry platform (Vitros 350) as the established method and two wet chemistry platforms (AU5800 and Cobas 6000) as the evaluation methods. Data thus generated was analyzed for any significant variation and tested if such variation increased with decreasing albumin/ globulin ratio.
RESULTS
Significant variation between dry chemistry and wet chemistry measurements were obtained for direct bilirubin, HDL and iron on AU5800 with p-values of 0.0009, <0.0001 and 0.0466 respectively. Similarly, discrepant results were obtained on Cobas 6000 for direct bilirubin and iron, with p-values of <0.0001 and 0.0002 respectively. Additionally, UIBC measurements on AU5800 varied significantly with increasing amounts of paraprotein present in the specimen (p-value = 0.0207).
CONCLUSION
This study emphasizes on predictive analyses to show that paraprotein interferences are fairly common on wet chemistry platforms. Evolving algorithms for monitoring of reaction curves can minimize release of erroneous results due to such interferences.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.