Endogenous monoclonal immunoglobulins analyzed using the EXENT® solution and LC-MS.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Barnidge David, Troske Derek, North Simon, Wallis Gregg, Perkins Mark, Harding Stephen

Аннотация

INTRODUCTION: The EXENT® Solution, a fully automated system, is a recent advancement for identifying and quantifying monoclonal immunoglobulins in serum. It combines immunoprecipitation with MALDI-TOF mass spectrometry. Compared to gel-based methods, like SPEP and IFE, it has demonstrated the ability to detect monoclonal immunoglobulins in serum at lower levels. In this study, samples that tested negative using EXENT® were reflexed to LC-MS to determine if the more sensitive LC-MS method could identify monoclonal immunoglobulins missed by EXENT®.

OBJECTIVES: To assess whether monoclonal immunoglobulins that are not detected by EXENT® can be detected by LC-MS using a low flow LC system coupled to a Q-TOF mass spectrometer.

METHODS: Samples obtained from patients confirmed to have multiple myeloma (MM) were diluted with pooled polyclonal human serum and analyzed using EXENT®. If a specific monoclonal immunoglobulin was not detected by EXENT®, the sample was then subjected to analysis by LC-MS. For the LC-MS analysis, the sample eluate, obtained after the MALDI-TOF MS spotting step, was collected and transferred to an autosampler tray for subsequent analysis using LC-MS.

CONCLUSION: LC-MS has the capability to detect monoclonal immunoglobulins that are no longer detected by EXENT®. Reflexing samples to LC-MS for analysis does not involve additional sample handling, allowing for a faster time-to-result compared to current approaches, such as Next-Generation Sequencing, Next-Generation Flow, and clonotypic peptide methods. Notably, LC-MS offers equivalent sensitivity in detecting these specific monoclonal immunoglobulins.