MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1371/journal.pone.0223597

Медицинская библиотека · PloS one · 2019

Can visual interpretation of NucliSens graphs reduce the need for repeat viral load testing?

Handireketi Newten, Timire Collins, Shewade Hemant Deepak, Munemo Ellen, Nyagupe Charles, Chipuka Sandra, Sisya Lucia, Gumbo Hlanai, Dhitima Ezekiel, Harries Anthony D

СтатьяПеревод в очереди

Аннотация

BACKGROUND

In Zimbabwe, viral load (VL) testing for people living with HIV on antiretroviral therapy is performed at the National Microbiology Reference Laboratory using a NucliSens machine. Anecdotal evidence has shown that invalid graphs for "Target Not Detectable (TND)" will upon repeat VL testing produce a valid result for virus not detected, therefore removing the need to repeat the test. This needs formal assessment.

OBJECTIVES

To determine i) intra- and inter-rater agreement of the visual interpretation of NucliSens graphs (Target Detectable [TD], TND and No Line [NL]) between two laboratory scientists and ii) sensitivity, specificity and predictive values of the NucliSens graphs compared with repeat VL results.

METHOD: Cross sectional study using secondary data. Two laboratory scientists independently rated graphs one week apart for intra-rater agreement and compared final ratings with each other for inter-rater agreement. Consensus interpretations of graphs were compared with repeat VL results. Kappa coefficients were used to obtain measures of agreement.

RESULTS

There were 562 patients with NucliSens graphs and repeat VL. Kappa scores were: 0.98 (Scientist A); 0.99 (Scientist B); 0.96 (Scientist A versus Scientist B); and 0.65 (NucliSens graphs versus VL). Sensitivity, specificity, positive predictive value and negative predictive value for graphs compared with VL were 71%, 92%, 79% and 89% respectively.

CONCLUSION

Intra-and inter-rater agreements were almost perfect. The negative predictive value translates to a false negative rate of 11%. If repeat VL testing is not done, the clinical consequences need to be balanced against cost savings and the risks outweigh the benefits.

Для цитирования: Handireketi Newten, Timire Collins, Shewade Hemant Deepak, Munemo Ellen, Nyagupe Charles, Chipuka Sandra, Sisya Lucia, Gumbo Hlanai, Dhitima Ezekiel, Harries Anthony D. Can visual interpretation of NucliSens graphs reduce the need for repeat viral load testing?. PloS one. 2019. DOI: 10.1371/journal.pone.0223597.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.