DOI 10.1016/0167-8140(95)01662-7

Медицинская библиотека · Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 1995

Changes to dose in the build-up region when using multi-leaf collimators in place of lead blocks supported on an accessory tray.

Adams E J, Hounsell A R

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

Doses in the build-up region have been compared for regular fields with and without a perspex accessory tray and for two irregular fields defined by either a Philips multi-leaf collimator (MLC) or lead blocks. The results show an increase in doses within the build-up region, by up to a factor of two, primarily due to the presence of the accessory tray. Fields shaped with the MLC exhibited build-up characteristics similar to those in unblocked fields and hence there will be systematic changes to the build-up dose when treatments are transferred from lead blocking techniques to those using an MLC mounted within the treatment head.

Для цитирования: Adams E J, Hounsell A R. Changes to dose in the build-up region when using multi-leaf collimators in place of lead blocks supported on an accessory tray.. Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology. 1995. DOI: 10.1016/0167-8140(95)01662-7.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.