Possible leakage radiation during malfunctioning of a Sagittaire accelerator.

Woudstra E, Huizenga H, van de Poel J A

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

A non-fatal accident with a CGR-Sagittaire accelerator in the Dr Daniel den Hoed Cancer Center (DDHCC) in Rotterdam in 1988 is described. In a period with frequently occurring technical problems, a patient, undergoing fractionated treatment at this accelerator for prostatic cancer, developed severe skin reactions on the right half ventral part of thorax, head and upper arm. The skin reactions suggested an irradiation with a dose of up to 10-20 Gy, so it was likely that a radiation accident had occurred caused by leakage radiation. A number of experiments were performed in attempting to explain the estimated large dose rates of leakage radiation. Under rather extreme malfunctioning conditions, a high leakage dose rate (4 Gy/min) could be obtained during therapy use. This condition might have occurred during this period of technical difficulty.

Для цитирования: Woudstra E, Huizenga H, van de Poel J A. Possible leakage radiation during malfunctioning of a Sagittaire accelerator.. Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology. 1993. DOI: 10.1016/0167-8140(93)90171-4.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.