Performance of a Veterinary Portable Glucose Meter for Blood Glucose Concentration Measurement in Red-Eared Slider Turtles (Trachemys scripta elegans).
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Cabble Ava, Liles Marina, Bonas Michael, Di Girolamo Nicola
Аннотация
BACKGROUND: There is limited knowledge about the accuracy and precision of portable blood glucose meters (PBGM) for measuring blood glucose in reptiles.
OBJECTIVE: We aimed to evaluate a veterinary PBGM for measuring blood glucose concentration in red-eared sliders by comparison with a laboratory reference method, and to assess the influence of packed cell volume on PBGM measurements.
METHODS: Blood samples were obtained from 20 red-eared sliders from the jugular vein. Two strips of a veterinary PBGM on the canine setting were run consecutively. Immediately afterward, the blood was centrifuged, and the plasma was separated into a clean tube. Plasma was frozen and shipped for glucose measurement via a hexokinase analyzer. Passing-Bablok regression and Bland-Altman analysis were conducted to determine agreement.
RESULTS: The veterinary PBGM overestimated blood glucose concentration, with a mean overestimation of 28.0 mg/dL (95% CI: 16.7-45.0, Limits of Agreement: 5.8-50.3 mg/dL). The bias was proportional, increasing at higher blood glucose concentrations. The mean difference was 23.1 mg/dL in the samples with PCV ≤ 19% (n = 9), 30.8 mg/dL in the samples with PCV between 20% and 29% (n = 7), and 34.4 mg/dL in the samples with PCV greater than or equal to 30% (n = 4). Intra-assay coefficients of variations, a measure of precision, were, on average, 4.5%, ranging from 0% to 17.4%.
CONCLUSIONS: The use of the veterinary PBGM evaluated in this study in red-eared sliders may alter both treatment and diagnostic decisions due to the overestimation of glucose concentrations.
OBJECTIVE: We aimed to evaluate a veterinary PBGM for measuring blood glucose concentration in red-eared sliders by comparison with a laboratory reference method, and to assess the influence of packed cell volume on PBGM measurements.
METHODS: Blood samples were obtained from 20 red-eared sliders from the jugular vein. Two strips of a veterinary PBGM on the canine setting were run consecutively. Immediately afterward, the blood was centrifuged, and the plasma was separated into a clean tube. Plasma was frozen and shipped for glucose measurement via a hexokinase analyzer. Passing-Bablok regression and Bland-Altman analysis were conducted to determine agreement.
RESULTS: The veterinary PBGM overestimated blood glucose concentration, with a mean overestimation of 28.0 mg/dL (95% CI: 16.7-45.0, Limits of Agreement: 5.8-50.3 mg/dL). The bias was proportional, increasing at higher blood glucose concentrations. The mean difference was 23.1 mg/dL in the samples with PCV ≤ 19% (n = 9), 30.8 mg/dL in the samples with PCV between 20% and 29% (n = 7), and 34.4 mg/dL in the samples with PCV greater than or equal to 30% (n = 4). Intra-assay coefficients of variations, a measure of precision, were, on average, 4.5%, ranging from 0% to 17.4%.
CONCLUSIONS: The use of the veterinary PBGM evaluated in this study in red-eared sliders may alter both treatment and diagnostic decisions due to the overestimation of glucose concentrations.