Treatment of postbiopsy pneumothorax with a self-contained pneumothorax treatment device.
СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
O Laoide R, Fundell L J, vanSonnenberg E, D'Agostino H, Oglevie S B, Rosenkrantz H
Аннотация
PURPOSE: To evaluate the usefulness of a self-contained pneumothorax treatment device, in which a catheter and a one-way valve compose a single unit, in cases of postbiopsy pneumothorax.
MATERIALS AND METHODS: Twenty-one patients underwent placement of the device to drain a postbiopsy pneumothorax. Treatment was considered successful when no therapy other than the device was needed.
RESULTS: Treatment was successful in 16 (76%) of the 21 patients. Causes of treatment failure were dislodgment of the device (n = 3), malfunction of the device (n = 1), and a persistent air leak that led to pneumothorax recurrence after removal of the device (n = 1). Eight patients (38%) experienced problems after insertion of the device: The device dislodged in seven patients and malfunctioned in one patient. These problems led to treatment failure in four patients (see above). In the remaining four, further treatment after device dislodgment was not required.
CONCLUSION: Treatment with the self-contained pneumothorax treatment device shows promise. Modifications to the device may alleviate technical problems and insertion difficulties.
MATERIALS AND METHODS: Twenty-one patients underwent placement of the device to drain a postbiopsy pneumothorax. Treatment was considered successful when no therapy other than the device was needed.
RESULTS: Treatment was successful in 16 (76%) of the 21 patients. Causes of treatment failure were dislodgment of the device (n = 3), malfunction of the device (n = 1), and a persistent air leak that led to pneumothorax recurrence after removal of the device (n = 1). Eight patients (38%) experienced problems after insertion of the device: The device dislodged in seven patients and malfunctioned in one patient. These problems led to treatment failure in four patients (see above). In the remaining four, further treatment after device dislodgment was not required.
CONCLUSION: Treatment with the self-contained pneumothorax treatment device shows promise. Modifications to the device may alleviate technical problems and insertion difficulties.