DOI 10.1007/s10840-021-00980-6
Superior vena cava isolation using a novel ablation catheter incorporating local impedance monitoring.
Аннотация
BACKGROUND
A novel technology able to measure the local impedance (LI) during radiofrequency ablation has become available for clinical use. We investigated the change in the LI characteristics during superior vena cava isolations (SVCIs) using a novel catheter equipped with mini-electrodes.
METHODS
Twenty paroxysmal atrial fibrillation patients (68 ± 9 years; 14 males) underwent an SVCI by targeting breakthroughs. Subsequently, dormant conduction provoked by adenosine triphosphate (ATP) was evaluated.
RESULTS
Electrical SVCIs were successfully achieved in all with 7.2 ± 3.0 radiofrequency applications (RFA) without any complications. The procedure and fluoroscopic times were 13.1 ± 8.1 and 2.8 ± 2.3 min. No ablation was required at the anteroseptal SVC in 19 (95.0%) patients. The baseline LI and generator impedance (GI) were 125 ± 23 and 105 ± 14Ω. LI drops during RFA were significantly greater than GI drops (17 ± 12 vs. 4 ± 4Ω, p < 0.001). The correlation between the LI drops and GI drops was relatively high (R = 0.69, p < 0.001). LI drops were highest at the septal SVC and lowest at the lateral followed by antero-lateral SVC. The baseline electrogram amplitude between the mini-electrodes and tip-ring electrodes was 1.2 ± 1.4 and 0.8 ± 0.6 mV. The mini-electrode amplitude is more sharply attenuated with a greater magnitude than the tip-ring amplitude (p < 0.001). ATP-provoked dormant conduction was exposed in 10/17 (58.8%) patients and antero-lateral SVC gap locations in 7. Antero-lateral SVC LI drops were similar between patients with and without dormancy.
CONCLUSIONS
The LI drop magnitude during RFA significantly differed among the SVC segments. Antero-lateral SVC ATP-provoked dormant conduction was often exposed, and additional applications are recommended following the isolation for a robust SVCI.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.