BACKGROUND

Late-onset complications of left atrial appendage occlusion (LAAO) device procedure are anecdotal and there are no such complications reported in literature using Cardia Ultraseal (Cardia, Inc., Eagan, MN, USA).

CASE SUMMARY: We report the case of a 74-year-old Caucasian man affected by paroxysmal atrial fibrillation with significant bleeding risk (familiar thrombocytopenia, macroscopic haematuria episodes during therapy with direct oral anticoagulants, HAS-BLED risk score: 4) and ischaemic risk as well (CHADSVASC score: 3). The patient was treated with LAAO device implantation for high bleeding risk. Subsequently, after 26 days from LAAO procedure, he was admitted to the emergency department for haematic cardiac tamponade. The patient was successfully treated with subxyphoidal pericardiocentesis in the acute phase, unfortunately cardiac arrest occurred during the transfer to the referral hospital for urgent cardiac surgery. Permanent neurological damage was reported and the patient died on day 28.

DISCUSSION: LAAO late-onset complications are very rare and the case presented is the first case described of late-onset pericardial effusion and tamponade secondary to the Cardia Ultraseal LAAO device implantation. We present a revision of the literature regarding the occurrence of similar adverse events and discuss the hypothetical mechanism of this major complication.

Для цитирования: Albani Stefano, Berlier Nicola, Pisano Francesco, Scacciatella Paolo. Left atrial appendage closure device complicated by late-onset pericardial effusion and tamponade: a case report.. European heart journal. Case reports. 2021. DOI: 10.1093/ehjcr/ytab058.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.