DOI 10.3390/diagnostics12030598

Медицинская библиотека · Diagnostics (Basel, Switzerland) · 2022

Comparison Study of the Two Biometers Based on Swept-Source Optical Coherence Tomography Technology.

Dong Jing, Yao Jinhan, Chang Shuimiao, Kanclerz Piotr, Khoramnia Ramin, Wang Xiaogang

СтатьяПеревод пока не подготовлен

Аннотация

This research aimed to investigate the potential differences in the parameters, including axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), lens thickness (LT), flat keratometry (Kf), steep keratometry (Ks), mean keratometry (Km), astigmatism, white-to-white (WTW) distance, acquired rate, and intraocular lens (IOL) power, between the two swept-source optical coherence tomography (SS-OCT) biometers, the ANTERION (biometer A) and IOLMaster 700 (biometer B). In a prospective observational comparative case series study, we enrolled 198 eyes undergoing cataract surgery. The AL, CCT, ACD, LT, Kf, Ks, Km, astigmatism, WTW, acquired rate, and IOL power were assessed. McNemar tests compared the acquired rate, and the paired sample t-test compared the quantitative measurement results between the groups. Nineteen eyes were excluded owing to missing AL data for either biometer. Finally, data from 179 eyes were analyzed. Between the two devices, no significant difference was found in AL, astigmatism magnitude, J0, and J45, while significant differences existed in CCT, ACD, LT, Kf, Ks, Km, WTW, astigmatism axis, and IOL power; no statistical significance was found in the AL acquired rate (biometer A, 90.9% and biometer B, 93.9%). Approximately 65.4% of eyes demonstrated ≥0.5-D difference in IOL power between the two biometers. In conclusion, the two biometers showed significant differences in all measurements (CCT, ACD, LT, K, WTW, astigmatism axis, and IOL power), except for AL.

Для цитирования: Dong Jing, Yao Jinhan, Chang Shuimiao, Kanclerz Piotr, Khoramnia Ramin, Wang Xiaogang. Comparison Study of the Two Biometers Based on Swept-Source Optical Coherence Tomography Technology.. Diagnostics (Basel, Switzerland). 2022. DOI: 10.3390/diagnostics12030598.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.