DOI 10.17116/oftalma20221380115
[Comparative assessment of the accuracy of toric intraocular lens calculations].
Аннотация
PURPOSE: To compare the accuracy of toric intraocular lens (IOL) calculations on three modern toric calculators.
MATERIAL AND
METHODS
The study comprised 35 eyes of 35 patients who underwent phacoemulsification with toric IOL implantation (EnVista Toric). Residual postoperative refractive astigmatism was calculated on three calculators: EnVista Toric Calculator, ASSORT Toric IOL Calculator and the Kane formula. Prediction error for each calculator was determined using vector analysis.
RESULTS
The mean absolute deviation of predicted postoperative refractive astigmatism over actual astigmatism in diopters was distributed in the following way: 0.82±0.58, 0.70±0.67 and 0.72±0.76 using EnVista Toric Calculator, ASSORT Toric IOL Calculator and the Kane formula, respectively. Centroid prediction error was 0.08 (EnVista Toric Calculator), 0.06 (ASSORT) and 0.10 (Kane formula). There was a significantly smaller deviation using ASSORT and the Kane formula compared to the online calculator (p<0.05).
CONCLUSIONS
Toric calculators ASSORT Toric IOL Calculator and the Kane formula showed higher accuracy of toric IOL calculation than EnVista Toric Calculator.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.