Super-large bore catheter in the treatment of large vessel occlusions: initial multicenter experience.
Аннотация
PURPOSE: Large bore catheters are increasingly used in mechanical thrombectomy (MT) for large vessel occlusions (LVOs).
OBJECTIVE
To evaluate the efficacy and safety of the super-large bore Cereglide 0.092" (C-92) catheter, featuring the largest inner diameter available.
METHODS
A multicenter observational study was conducted across 12 comprehensive stroke centers in the United States. Efficacy outcomes included the first pass effect (FPE) and successful reperfusion. FPE was defined as a first MT pass achieving a modified Treatment in Cerebral Infarction (mTICI) score of ≥2c. Successful reperfusion was defined as final mTICI score ≥2c. Safety outcomes involved device-related complications, symptomatic intracranial hemorrhage (sICH), and inpatient mortality. Functional outcomes included modified Rankin Scale (mRS) score at discharge and delta National Institutes of Health Stroke Scale (NIHSS) score.
RESULTS
Fifty patients were included. The most common LVO was the first segment of the middle cerebral artery in 31/50 cases (62%). The C-92 reached the thrombus in 41 patients (82%). Median puncture-to-thrombus and puncture-to-reperfusion times were 15 min (IQR 10-25) and 26 min (IQR 15-49), respectively. FPE was achieved in 25/50 (50%) cases, and in 25/41 (61%) cases when the C-92 reached the thrombus. Successful reperfusion occurred in 36/41 patients (88%). There were no vessel perforations, or sICH. Distal embolization occurred in 4/50 (8%) cases, and 4/50 (8%) died. The mRS score at discharge was 3 (IQR 2-6), and the delta NIHSS score was 8 (IQR 5-12).
CONCLUSION
The C-92 catheter demonstrated a safe profile achieving an overall FPE rate of 50%, and favorable functional outcomes in 88% of cases.