Biometric and intraocular pressure change after cataract surgery.

Bilak Semsettin, Simsek Ali, Capkin Musa, Guler Mete, Bilgin Burak

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate changes in ocular biometric parameters with optical biometry and intraocular pressure (IOP) by Goldmann tonometry after uneventful phacoemulsification in normotensive eyes.

METHODS

This prospective study was composed of 117 eyes of 117 patients who had undergone uncomplicated phacoemulsification and foldable intraocular lens implantation. The measurements were obtained preoperatively and 1 month postoperatively. Axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), pupil diameter (PD), and lens thickness were measured by the LenStar optical biometer, and IOP was measured by Goldmann applanation tonometry. The relationships between IOP change and ocular biometric parameters were evaluated.

RESULTS

Mean age was 64.73 ± 10.83 years; 45 patients were women (38.46%) and 72 patients were men (61.54%). Comparing preoperative and postoperative 1-month measurements, the mean AL, PD, and IOP decreased (p < 0.01), whereas ACD (p < 0.01) and CCT (p > 0.05) increased. Intraocular pressure change correlated positively with AL change (p < 0.05) and PD change (p < 0.05) and preoperative IOP (p < 0.01) and preoperative CCT (p < 0.05). Anterior chamber depth change correlated positively with preoperative ACD (p < 0.01) and negatively with preoperative PD (p < 0.05). Pupil diameter change positively correlated with preoperative PD (p < 0.01). Regression analyses showed that preoperative IOP significantly associated with postoperative IOP reduction according to the standardized coefficient beta (-0.649).

CONCLUSIONS

Anterior segment morphometry changed and IOP decreased after phacoemulsification. The preoperative IOP may be the best parameter for estimating postoperative IOP reduction.

Для цитирования: Bilak Semsettin, Simsek Ali, Capkin Musa, Guler Mete, Bilgin Burak. Biometric and intraocular pressure change after cataract surgery.. Optometry and vision science : official publication of the American Academy of Optometry. 2015. DOI: 10.1097/opx.0000000000000553.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.