Harrer Annette, Hirnschall Nino, Maedel Sophie, Findl Oliver
Аннотация
PURPOSE: To evaluate the rotational stability of two intraocular lenses (IOLs) of similar design and material but with a difference of 1 mm in overall length.
METHODS: In this prospective study patients with age-related cataract were included. An IOL with an overall diameter of 12 mm (ACR6 = small-diameter IOL) was compared to an IOL with an overall diameter of 13 mm (IDEA 613 XC = large-diameter IOL).
RESULTS: In total, 60 patients were included in this study. Absolute rotation in the small- and large-diameter groups was 4.4° (SD: 4.0; range: 0.3-17.8) and 3.0° (SD: 2.4; range: 0.1-7.8), respectively. The differences between the two IOLs were not found to be statistically significant.
CONCLUSION: The effect of the overall length of an IOL appears to have little impact on early rotation after cataract surgery.
METHODS: In this prospective study patients with age-related cataract were included. An IOL with an overall diameter of 12 mm (ACR6 = small-diameter IOL) was compared to an IOL with an overall diameter of 13 mm (IDEA 613 XC = large-diameter IOL).
RESULTS: In total, 60 patients were included in this study. Absolute rotation in the small- and large-diameter groups was 4.4° (SD: 4.0; range: 0.3-17.8) and 3.0° (SD: 2.4; range: 0.1-7.8), respectively. The differences between the two IOLs were not found to be statistically significant.
CONCLUSION: The effect of the overall length of an IOL appears to have little impact on early rotation after cataract surgery.