DOI 10.3109/02713683.2014.968936

Медицинская библиотека · Current eye research · 2014

Prospective Comparison of One-Piece and Three-Piece Tecnis Aspheric Intraocular Lenses: 1-year Stability and its Effect on Visual Function.

Miyata Kazunori, Kataoka Yasushi, Matsunaga Jiro, Honbo Masato, Minami Keiichiro

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To compare the stability and visual function of one-piece hydrophobic intraocular lens (IOL) (ZCB00, Abbot Medical Optics) and a three-piece hydrophobic IOL (ZA9003, Abbot Medical Optics) after cataract surgery.

MATERIALS AND

METHODS

Eighteen eyes of 18 patients with the one-piece IOL and 20 eyes of 20 patients with the three-piece IOL who underwent cataract surgery were evaluated prospectively on best corrected distance visual acuity, postoperative refractive error, IOL decentration and tilt, anterior chamber depth (ACD), spherical aberration and contrast sensitivity. Measurements were performed 1 week and 1, 3, 6 and 12 months postoperatively. The postoperative outcomes were compared between the two groups.

RESULTS

In the one-piece group, a hyperopic refractive error (p < 0.002) and deeper ACDs during 1-year post-operatively were observed compared with the three-piece group. There were no significant differences in BCVA, IOL tilt, spherical aberration or contrast sensitivity.

CONCLUSIONS

The stability of the one-piece IOL was greater than that of the three-piece IOL. The postoperative ACD and refractive error demonstrated that an anteriorly offset haptic in the one-piece IOL allowed incremental fixation to the posterior capsule.

Для цитирования: Miyata Kazunori, Kataoka Yasushi, Matsunaga Jiro, Honbo Masato, Minami Keiichiro. Prospective Comparison of One-Piece and Three-Piece Tecnis Aspheric Intraocular Lenses: 1-year Stability and its Effect on Visual Function.. Current eye research. 2014. DOI: 10.3109/02713683.2014.968936.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.