DOI 10.1016/j.jcrs.2013.10.035

Медицинская библиотека · Journal of cataract and refractive surgery · 2014

Comparison of methods to quantify posterior capsule opacification using forward and backward light scattering.

Hirnschall Nino, Crnej Alja, Gangwani Vinod, Findl Oliver

Сравнительное исследованиеСтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To compare forward and backward light scattering measurements to quantify posterior capsule opacification (PCO).

SETTING

Vienna Institute for Research in Ocular Surgery, Vienna, Austria.

DESIGN

Prospective single-center study.

METHODS

This observational study comprised consecutive patients scheduled for neodymium:YAG (Nd:YAG) capsulotomy for regeneratory PCO. The corrected distance visual acuity (CDVA) using Early Treatment of Diabetic Retinopathy Study charts and the contrast sensitivity under mesopic and photopic conditions using Pelli-Robson charts were assessed before an Nd:YAG capsulotomy was performed. Retroillumination images, rotating Scheimpflug scans (Pentacam HR), and straylight meter (C-Quant) and point-spread function (PSF) meter (Optical Quality Analysis System) measurements were performed before and after Nd:YAG capsulotomy.

RESULTS

The study enrolled 50 eyes of 50 patients. The mean uncorrected distance visual acuity and CDVA were 0.76 logMAR ± 0.18 (SD) and 0.68 ± 0.2 logMAR, respectively. The mean log contrast sensitivity (logCS) was 1.2 ± 0.15 logCS under photopic conditions and 1.05 ± 0.15 logCS under mesopic conditions. There was a moderate correlation between the Scheimpflug score and the Automated Quantification of After-Cataract score (r(2) = 0.37, P=.03). Furthermore, a moderate and close to significant correlation between the PSF meter and the straylight meter was observed (r(2) = 0.32; P=.07).

CONCLUSION

Each of the 4 devices assesses different aspects of a patient's quality of vision. Further developments should focus on methods that measure forward scatter of light in a large visual angle and are not too patient or examiner dependent.

Для цитирования: Hirnschall Nino, Crnej Alja, Gangwani Vinod, Findl Oliver. Comparison of methods to quantify posterior capsule opacification using forward and backward light scattering.. Journal of cataract and refractive surgery. 2014. DOI: 10.1016/j.jcrs.2013.10.035.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.