Optometry and vision science : official publication of the American Academy of Optometry · 2014
Double-pass system assessing the optical quality of pseudophakic eyes.
Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Lee Hun, Lee Kwanghyun, Ahn Ji Min, Kim Eung Kweon, Sgrignoli Bradford, Kim Tae-Im
Аннотация
PURPOSE: To compare the optical quality measurements obtained from the double-pass system and ocular aberrations, subjective visual acuity, and contrast sensitivity score in pseudophakic eyes.
METHODS: Three months after cataract surgery, modulation transfer function (MTF) cutoff frequency, Strehl ratio, objective scatter index, and objective pseudoaccommodation obtained from the double-pass system were compared with total aberration, higher-order aberration, and spherical aberration obtained from ray-tracing aberrometer. In addition, parameters of the double-pass system were compared with subjective visual acuity and the contrast sensitivity score.
RESULTS: Forty eyes of 40 patients were included. The MTF cutoff frequency and Strehl ratio were negatively correlated with total aberration (r = -0.503, p = 0.003; r = -0.509, p = 0.003, respectively) and subjective visual acuity (r = -0.453, p = 0.007; r = -0.354, p = 0.040, respectively). The objective scatter index was positively correlated with total aberration (r = 0.451, p = 0.024) and subjective visual acuity (r = 0.516, p = 0.008). The MTF cutoff frequency showed a correlation with contrast sensitivity score under photopic and mesopic conditions.
CONCLUSIONS: Optical quality parameters obtained from the double-pass system were correlated with ocular aberrations, subjective visual acuity, and contrast sensitivity score in pseudophakic eyes.
METHODS: Three months after cataract surgery, modulation transfer function (MTF) cutoff frequency, Strehl ratio, objective scatter index, and objective pseudoaccommodation obtained from the double-pass system were compared with total aberration, higher-order aberration, and spherical aberration obtained from ray-tracing aberrometer. In addition, parameters of the double-pass system were compared with subjective visual acuity and the contrast sensitivity score.
RESULTS: Forty eyes of 40 patients were included. The MTF cutoff frequency and Strehl ratio were negatively correlated with total aberration (r = -0.503, p = 0.003; r = -0.509, p = 0.003, respectively) and subjective visual acuity (r = -0.453, p = 0.007; r = -0.354, p = 0.040, respectively). The objective scatter index was positively correlated with total aberration (r = 0.451, p = 0.024) and subjective visual acuity (r = 0.516, p = 0.008). The MTF cutoff frequency showed a correlation with contrast sensitivity score under photopic and mesopic conditions.
CONCLUSIONS: Optical quality parameters obtained from the double-pass system were correlated with ocular aberrations, subjective visual acuity, and contrast sensitivity score in pseudophakic eyes.