Jornal de pediatria · 2023

Comparison of the effectiveness of different high-flow devices in neonatal care.

Наблюдательное исследованиеСтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Kaminski Diogo Machado, Biazus Graziela Ferreira, Silveira Rita C, Procianoy Renato Soibelmann

Аннотация

OBJECTIVE: To evaluate the success rate of high-flow nasal cannula (HFNC) therapy using an adapted obsolete mechanical ventilator (MV), Optiflow™ and Vapotherm™ in newborns (NBs).

METHOD: This was a retrospective observational study conducted in the neonatal intensive care unit (NICU). The sample comprised NBs who underwent HFNC therapy due to ventilatory dysfunction, for weaning from non-invasive ventilation (NIV), or post-extubation. The three groups, stratified according to gestational age (GA) and birth weight, and corrected GA and weight at the beginning of HFNC use, were as follows: Optiflow ™, Vapotherm ™, and obsolete Mechanical Ventilator (MV) adapted for high flow therapy. Subsequently, the NBs were divided into a success group (SG) and a failure group (FG). HFNC success was defined as a therapy duration exceeding 72 h.

RESULTS: A total of 209 NBs were evaluated, with 31.1 % using HFNC due to ventilatory dysfunction, 2.4 % after extubation, and 66.5 % after NIV weaning. HFNC success rate was observed in 90.9 % of the NBs, with no difference between equipment types (Vapotherm ™, Optiflow ™, and adapted VM).

CONCLUSION: Different types of HFNC equipment are equally effective when used in neonatology for respiratory dysfunction, as a method of weaning from NIV and post-extubation. Adapted obsolete MV can be an alternative for HFCN therapy in resource-constrained settings.