m+Medelevator.RU
← Поиск публикаций
Archivos de la Sociedad Espanola de Oftalmologia · 2025

Comparison of 2 swept-source optical biometers: IOLMaster 700 vs. Revo FC 130.

СтатьяНаблюдательное исследованиеСравнительное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Ortiz A, García E, Ortega J, de Vera J, Vázquez E, Carrillo M, Romero V, Merizalde K

Аннотация

OBJECTIVE: To evaluate the measurement accuracy of 2 swept-source optical biometers, the IOLMaster 700 and the Revo FC 130, in patients diagnosed with cataracts.

METHODS: We conducted a prospective, observational and comparative study in patients older than 40 years with cataracts. A single operator performed optical biometry using both devices during the same session. Cataract grading was conducted using the Lens Opacities Classification System III.

RESULTS: Biometric measurements were obtained from the 504 eyes of 279 patients, with a mean age of 70.93 ± 8.41 years. Statistically significant differences were found between the IOLMaster 700 and Revo FC 130 in the following parameters: mean keratometry (43.53 ± 1.56 vs 44.27 ± 1.68 diopters), total average keratometry and mean keratometry (43.56 ± 1.55 D vs 44.27 ± 1.67 diopters), white-to-white distance (11.98 ± 0.52 vs 12.66 ± 0.73 mm), and lens thickness (4.53 ± 0.40 vs 4.60 ± 0.40 mm). No significant differences were found in axial length (23.31 ± 0.96 vs 23.25 ± 1.01 mm) or anterior chamber depth (3.02 ± 0.38 vs 3.02 ± 0.37 mm), which were not clinically relevant. No statistically significant differences were found between the IOLMaster 700 measurements of total mean keratometry (43.53 ± 1.56 diopters) and mean keratometry (43.56 ± 1.56 diopters).

CONCLUSIONS: Although most measurements do not reveal any clinically relevant differences, the 2 devices should not be used interchangeably, mainly due to variations in keratometry readings.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты