Comparison of 2 swept-source optical biometers: IOLMaster 700 vs. Revo FC 130.

Ortiz A, García E, Ortega J, de Vera J, Vázquez E, Carrillo M, Romero V, Merizalde K

СтатьяНаблюдательное исследованиеСравнительное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVE

To evaluate the measurement accuracy of 2 swept-source optical biometers, the IOLMaster 700 and the Revo FC 130, in patients diagnosed with cataracts.

METHODS

We conducted a prospective, observational and comparative study in patients older than 40 years with cataracts. A single operator performed optical biometry using both devices during the same session. Cataract grading was conducted using the Lens Opacities Classification System III.

RESULTS

Biometric measurements were obtained from the 504 eyes of 279 patients, with a mean age of 70.93 ± 8.41 years. Statistically significant differences were found between the IOLMaster 700 and Revo FC 130 in the following parameters: mean keratometry (43.53 ± 1.56 vs 44.27 ± 1.68 diopters), total average keratometry and mean keratometry (43.56 ± 1.55 D vs 44.27 ± 1.67 diopters), white-to-white distance (11.98 ± 0.52 vs 12.66 ± 0.73 mm), and lens thickness (4.53 ± 0.40 vs 4.60 ± 0.40 mm). No significant differences were found in axial length (23.31 ± 0.96 vs 23.25 ± 1.01 mm) or anterior chamber depth (3.02 ± 0.38 vs 3.02 ± 0.37 mm), which were not clinically relevant. No statistically significant differences were found between the IOLMaster 700 measurements of total mean keratometry (43.53 ± 1.56 diopters) and mean keratometry (43.56 ± 1.56 diopters).

CONCLUSIONS

Although most measurements do not reveal any clinically relevant differences, the 2 devices should not be used interchangeably, mainly due to variations in keratometry readings.

Для цитирования: Ortiz A, García E, Ortega J, de Vera J, Vázquez E, Carrillo M, Romero V, Merizalde K. Comparison of 2 swept-source optical biometers: IOLMaster 700 vs. Revo FC 130.. Archivos de la Sociedad Espanola de Oftalmologia. 2025. DOI: 10.1016/j.oftale.2025.09.002.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.