m+Medelevator.RU
← Поиск публикаций
Journal of cataract and refractive surgery · 2025

Effect of suction angle on zonular fibers during residual cortex removal in phacoemulsification using finite element modeling.

СтатьяНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Daryabari Seyed Hashem, Hosseini Imeni Seyyed Morteza, Ghouhestani Mojtaba

Аннотация

PURPOSE: To evaluate the effect of suction direction on zonular fibers' mechanical behavior using finite element modeling.

SETTING: Laboratory study.

DESIGN: 3D finite element model experimental study.

METHODS: A 3D finite element model was developed, including nucleus covered by cortex and capsule. Similar to cataract surgery, a circular rapture was considered at the top of the capsule to apply suction pressure. Finally, zonular fibers were modeled as a continuum body using 3D solid elements (C3D8R). A custom FORTRAN subroutine was implemented to enforce tension-only behavior, mimicking the physiological characteristics of the zonular fibers, which are resistant to tensile but not compressive loads. This method allows for a realistic simulation of zonular mechanics during cortical aspiration.

RESULTS: The suction processes in tangential and normal angles were simulated. The resulting relative displacements between the cortex and capsule, as a criterion of cortex separation, vs the resulting maximum zonule displacements, were recorded in each model. The cortex-capsule relative displacement vs maximum zonule displacement indicated a diagram slope of 0.09 for tangential applying pressure and 0.02 in the case of applying normal pressure.

CONCLUSIONS: The results illustrated that zonular fibers were less tensile in a specific magnitude of cortex-capsule relative displacement under tangential applying pressure, which means that eye zonular fibers have a lower risk of failure until the separation of the lens cortex.

Предметные рубрики MeSH

ИП Урсолов Владлен Сергеевич

ИНН 500118482352 · ОГРНИП 314500117700049

Реквизиты и контакты