Vascular medicine (London, England) · 2025

Relationship between exercise transcutaneous oxygen pressure measurements with constant and graded treadmill tests.

Сравнительное исследованиеСтатьяНаблюдательное исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Vannier Cindy, Guézais Adélaïde, Talbot Marjolaine, Le Faucheur Alexis, Mahé Guillaume

Аннотация

INTRODUCTION: Exercise transcutaneous oxygen pressure measurement (Exercise-TcPO2) that has been validated with a constant-load treadmill protocol, is useful to assess patients with suspected lower-extremity peripheral artery disease (PAD). To date, nothing is known about its use with a graded protocol. The aim of this study is to compare the Exercise-TcPO2 results obtained in patients with suspected PAD using a graded Gardner-Skinner (3.2 km/h, 2% grade increase every 2 min) versus a constant modified Strandness (3.2 km/h, 10% grade) treadmill test.

METHODS: Patients with suspected PAD were assessed twice on the same day on a treadmill using the modified Strandness (Strand) followed by the Gardner-Skinner (Gard) tests. In both tests, the decrease from resting oxygen pressure (DROP) indexes were measured at the buttock and calf levels as previously validated. A DROPStrand index ⩽ -15 mmHg was considered positive for the diagnosis of PAD. The relationship between DROPGard and DROPStrand as well as DROP value differences between the two tests and concordances were studied.

RESULTS: This prospective study included 40 patients. Using a linear mixed model, the relationship between DROPGard and DROPStrand was defined as DROPGard = 0.647 × DROPStrand - 1.858. The mean difference between DROPGard and DROPStrand was -4.3 ± 7.3 mmHg. Positive and negative agreements were 78.0% and 86.3%, respectively. The kappa coefficient between DROPGard and DROPStrand was 0.65 [0.54; 0.77].

CONCLUSION: Exercise-TcPO2 results using the Gardner-Skinner and modified Strandness treadmill tests were highly correlated. This study suggests the use of a DROPGard value ⩽ -12 mmHg to diagnose PAD when performing Exercise-TcPO2 with a Gardner-Skinner test. However, this cut-off value should be confirmed against an imaging gold standard. ClinicalTrials.gov identifier: NCT03186391.