[The ability of frequency doubling technology to detect abnormality of visual function in early glaucoma].

Yamashiro H, Tanaka M, Saito M, Shirato S

Аннотация на английскомСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate the ability of the Frequency Doubling Technology(FDT) threshold test in detecting abnormality of visual function in early glaucoma patients.

SUBJECTS AND METHOD: The C-20 full threshold test of FDT was performed on 34 normal-tension glaucoma(NTG) patients and 39 primary open-angle glaucoma(POAG) patients with visual field defects limited to the upper or lower hemi-field as detected by Humphrey Field Analyzer(HFA) and on 79 normal control subjects. Optic disk findings and FDT results corresponding to the intact hemi-field were evaluated. FDT abnormalities in normal subjects were calculated as false positive rates in FDT.

RESULT: The sensitivity and specificity of FDT, calculated based on optic disk findings, were 75.0% and 61.1% in POAG, and 61.1% and 66.7% in NTG, respectively, while the false positive rate in normal subjects was 11.4%.

CONCLUSION

FDT can detect glaucomatous functional abnormality earlier than HFA.

Для цитирования: Yamashiro H, Tanaka M, Saito M, Shirato S. [The ability of frequency doubling technology to detect abnormality of visual function in early glaucoma].. Nippon Ganka Gakkai zasshi. 2001.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.