Pupil perimetry using M-sequence stimulation technique.
Аннотация
PURPOSE: M-sequence stimulation technique allows mapping of the retinal function by multifocal electroretinographic (ERG) recordings. However, the information provided about visual field is limited to retinal function. Optic nerve diseases and diseases of the higher visual pathways usually show normal multifocal ERGs. Using pupillary responses instead of the electrical retinal responses might enhance the diagnostic possibilities of this system. The problems of local ERG recordings are very similar to those encountered in pupil perimetry: Local stimuli have to be dim to avoid or at least reduce stray-light responses. Dim stimuli, close to the absolute threshold, elicit only subtle pupillomotor responses. Therefore, techniques that are able to detect small focal responses are promising.
METHODS
Pupillography was done by means of an infrared video camera and real time image processing (50 Hz) using a custom-designed videoboard in a personal computer (486). Recording conditions: The stimulus was presented on a monitor (75 Hz) in 26 cm distance from the patient's eyes. It contained 37 hexagons in a 25 degrees visual field. Each element changed between black (1.6 cd/m2) and white (160 cd/m2) after a binary M-sequence independently from other elements. Four thousand ninety six different stimulus pictures of 120-msec duration were shown during a single pupillogram recording. Thirty-seven local pupillograms were calculated in a cross-correlation of stimulus sequence and the pupil diameter.
RESULTS
The pupillomotor fields in normals showed a shape and sensitivity distribution as known from conventional pupil perimetry techniques. Artificial paracentral scotomas (5 degrees) created by masking different locations could be demonstrated convincingly. Even in patients with optic nerve lesions it was possible to demonstrate visual field defects.
CONCLUSIONS
Pupil perimetry using the M-sequence technique is a promising method of objective perimetry that may find its entrance into clinical application.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.