DOI 10.1177/19322968251315459

Медицинская библиотека · Journal of diabetes science and technology · 2025

Performance of Three Continuous Glucose Monitoring Systems in Adults With Type 1 Diabetes.

Eichenlaub Manuel, Waldenmaier Delia, Wehrstedt Stephanie, Pleus Stefan, Link Manuela, Jendrike Nina, Öter Sükrü, Haug Cornelia, Schinz Maren, Braunack-Mayer Vincent, Schneider Regula, Brandt Derek, Freckmann Guido

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

The performance of continuous glucose monitoring (CGM) systems is difficult to compare due to different study designs and a lack of head-to-head studies. This study evaluated the performance of FreeStyle Libre 3 (FL3), Dexcom G7 (DG7), and Medtronic Simplera (MSP) against different comparator methods and during clinically relevant glycemic scenarios.

METHOD: Twenty-four adult participants with type 1 diabetes mellitus wore one sensor of each CGM system in parallel for up to 15 days. Sensors of DG7 and MSP were exchanged on days 5 and 8, respectively. Three 7-hour sessions with 15-minute comparator blood glucose-level measurements using YSI 2300 (YSI, venous), Cobas Integra (INT, venous), and Contour Next (CNX, capillary) were conducted on days 2, 5, and 15. Simultaneously, glucose-level excursions with transient hyperglycemia and hypoglycemia were induced according to a recently published testing procedure. The accuracy was evaluated using various metrics, including mean absolute relative differences (MARDs).

RESULTS

Compared with YSI data, the MARDs of FL3, DG7, and MSP were 11.6%, 12.0%, and 11.6%, respectively. Relative to the INT data, the corresponding MARDs were 9.5%, 9.9%, and 13.9%, respectively, and compared with CNX data, MARDs were 9.7%, 10.1%, and 16.6%, respectively. Both FL3 and DG7 showed better accuracy in the normoglycemic and hyperglycemic range, while MSP performed better in the hypoglycemic range.

CONCLUSIONS

Performance results of all CGM systems varied depending on the comparator method. However, across comparators FL3 and DG7 tended to be more accurate compared with MSP. All CGM systems showed a lower accuracy compared with previous studies, emphasizing the need for comprehensive study design guidelines.

Для цитирования: Eichenlaub Manuel, Waldenmaier Delia, Wehrstedt Stephanie, Pleus Stefan, Link Manuela, Jendrike Nina, Öter Sükrü, Haug Cornelia, Schinz Maren, Braunack-Mayer Vincent, Schneider Regula, Brandt Derek, Freckmann Guido. Performance of Three Continuous Glucose Monitoring Systems in Adults With Type 1 Diabetes.. Journal of diabetes science and technology. 2025. DOI: 10.1177/19322968251315459.