Performance of Three Continuous Glucose Monitoring Systems in Adults With Type 1 Diabetes.
Аннотация
BACKGROUND
The performance of continuous glucose monitoring (CGM) systems is difficult to compare due to different study designs and a lack of head-to-head studies. This study evaluated the performance of FreeStyle Libre 3 (FL3), Dexcom G7 (DG7), and Medtronic Simplera (MSP) against different comparator methods and during clinically relevant glycemic scenarios.
METHOD: Twenty-four adult participants with type 1 diabetes mellitus wore one sensor of each CGM system in parallel for up to 15 days. Sensors of DG7 and MSP were exchanged on days 5 and 8, respectively. Three 7-hour sessions with 15-minute comparator blood glucose-level measurements using YSI 2300 (YSI, venous), Cobas Integra (INT, venous), and Contour Next (CNX, capillary) were conducted on days 2, 5, and 15. Simultaneously, glucose-level excursions with transient hyperglycemia and hypoglycemia were induced according to a recently published testing procedure. The accuracy was evaluated using various metrics, including mean absolute relative differences (MARDs).
RESULTS
Compared with YSI data, the MARDs of FL3, DG7, and MSP were 11.6%, 12.0%, and 11.6%, respectively. Relative to the INT data, the corresponding MARDs were 9.5%, 9.9%, and 13.9%, respectively, and compared with CNX data, MARDs were 9.7%, 10.1%, and 16.6%, respectively. Both FL3 and DG7 showed better accuracy in the normoglycemic and hyperglycemic range, while MSP performed better in the hypoglycemic range.
CONCLUSIONS
Performance results of all CGM systems varied depending on the comparator method. However, across comparators FL3 and DG7 tended to be more accurate compared with MSP. All CGM systems showed a lower accuracy compared with previous studies, emphasizing the need for comprehensive study design guidelines.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.