Pacing failure due to microdislodgement of ventricular pacing lead detected by home monitoring technology.

Loricchio Maria Luisa, Castro Antonio, Ciolli Andrea, Sasdelli Massimo, Ferraiuolo Giuseppe

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

A 68-year-old woman affected by sick sinus syndrome was implanted with a dual-chamber pacemaker provided by home monitoring technology. After discharge, an increase in ventricular threshold and a high variability of R wave measurements were detected early by the home monitoring system. Manual tests confirmed the presence of pacing and sensing failure and a normal ventricular impedance. The pacing lead integrity and a stable position of the lead tip in right ventricular apex were assessed by chest X-ray. A diagnosis of microdislodgement was made. After a second procedure for ventricular lead repositioning, no further malfunctions were detected.

Для цитирования: Loricchio Maria Luisa, Castro Antonio, Ciolli Andrea, Sasdelli Massimo, Ferraiuolo Giuseppe. Pacing failure due to microdislodgement of ventricular pacing lead detected by home monitoring technology.. Journal of cardiovascular medicine (Hagerstown, Md.). 2008. DOI: 10.2459/jcm.0b013e328301355f.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.